3 - 11
3.7.2
Privacy Mode
The privacy mode is a special clinical monitoring mode. In the privacy mode, the monitor does not display
patient information and monitoring data. This provides controlled access to patient data and ensures
confidentiality.
The privacy mode is only available when the patient admitted by the monitor is also monitored by the CMS. The
monitor continues monitoring the patient, but patient data is only visible at the CMS.
3.7.2.1
Entering the Privacy Mode
To enter the privacy mode, choose either of the following ways:
■
Select the
Privacy Mode
quick key
→
select
OK
.
■
Select the
Main Menu
quick key
→
from the
Display
column select
Privacy Mode
→
select
OK
.
The monitor has the following features after entering the privacy mode:
■
The screen turns blank.
■
Except for the low battery alarm, the monitor inactivate alarm tone and alarm light of all other alarms.
■
The monitor suppresses all system sounds, including heart beat tone, pulse tone, and prompt tone.
WARNING
.
•
In Privacy mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor.
Alarms are presented only at the CMS. Pay attention to potential risk.
NOTE
•
The privacy mode is not available if the Department is set to OR.
•
You cannot enter the privacy mode if a low battery alarm occurs.
3.7.2.2
Exiting the Privacy Mode
The monitor automatically exit the privacy mode in any of the following situations:
■
The monitor disconnects from the CMS.
■
The low battery alarm occurs.
You can also operate the touchscreen to manually exit the privacy mode.
3.7.3
Night Mode
The night mode is a special clinical monitoring mode. To avoid disturbing the patient, you can use the night
mode.
3.7.3.1
Entering the Night Mode
To enter the night mode, follow this procedure:
1.
Select the
Night Mode
quick key, or select the
Main Menu
quick key
→
from the
Display
column select
Night Mode
.
2.
Change the night mode settings if necessary.
3.
Select
Enter Night Mode
.
The night mode settings are as follows by default:
■
Brightness
:
1
■
Alarm Volume
:
2
■
QRS Volume
:
1
■
Key Volume
:
0
■
NIBP End Tone
:
Off
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...