E - 3
For BF
applied parts
◆
100 μA in Normal Condition
◆
500 μA in Single Fault Condition
E.7
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a limiting
resistance, to selected applied part terminals. Current measurements are then taken between the selected
applied part and earth. Measurements are taken with the test voltage (110% of mains) to applied parts in the
normal and reverse polarity conditions
The following outlet conditions apply when performing the Mains on Applied Part test.
■
Normal Polarity;
■
Reversed Polarity
LIMITS
◆
For CF
applied parts: 50 μA
◆
For BF
applied parts: 5000 μA
E.8
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining
Applied Part connector s. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
■
normal polarity (Normal Condition);
■
reverse polarity (Normal Condition),
■
normal polarity with open neutral (Single Fault Condition);
■
reverse polarity with open neutral (Single Fault Condition).
■
normal polarity with open earth (Single Fault Condition);
■
reverse polarity with open earth (Single Fault Condition).
LIMITS
For CF
applied parts,
◆
10 μA in Normal Condition
◆
50 μA in Single Fault Condition
For BF
applied parts,
◆
100 μA in Normal Condition
◆
500 μA in Single Fault Condition
NOTE
•
Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
•
Follow the instructions of the analyzer manufacturer.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...