25 - 3
25.4
Automatically Printing Reports
When a parameter alarm switch is set to on and an alarm is triggered for this parameter, you can set a printer to
start alarm printing automatically.
To do so, follow this procedure:
1.
Access alarm related tabs such as the
Alarm
tab for a parameter in one of the following ways:
◆
Select the
Alarm Setup
quick key at the bottom of the screen.
◆
Select the parameter or waveform area of the desired parameter
→
select the
Alarm
tab.
◆
Select the
Parameters Setup
quick key at the bottom of the screen
→
select the desired parameter
→
select the
Alarm
tab.
2.
Switch on
Alarm Outputs
for desired parameters.
25.5
Stopping a Printing Task
To stop a printing task, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Report
column select
Print Queue
.
2.
Select desired printing tasks and then select
Delete
. Selecting
Delete All
to stop all the printing tasks.
25.6
Setting Reports
This section focuses on how to set ECG reports, realtime reports, tabular trends reports, and graphic trends
reports.
25.6.1
Setting ECG Reports
To set ECG reports, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Report
column select
Normal Report
.
2.
Select
ECG Report
.
3.
Set the desired options.
The following table only lists some of the options.
Menu item
Function
Description
Speed
Set the print speed of
ECG waveforms
25 mm/sec: prints 25 mm of ECG waveform per second.
50 mm/sec: prints 50 mm of ECG waveform per second.
Auto Interval
Defines the spacing
between the ECG
waveforms
on a
printout
On:
automatically adjusts the space between waveforms to avoid
overlapping.
Off
: each waveform area has the same size on a printout.
Note: This setting is only relevant when
12×1
is selected for
12-Lead Format
.
12-Lead Format
Select the format of 12-
lead ECG waveforms
on a printout.
12×1:
displays 12-lead ECG waveforms on one page in one column.
6×2
: displays 12-lead ECG waveforms on one page in two columns,
with 6 lines in each column.
6×2+1
: displays 12-lead ECG waveforms on one page in two
columns, with 6 lines in each column, and one rhythm lead
waveform at the bottom.
3×4+1
: displays 12-lead ECG waveforms on one page in 4 columns,
with 3 lines in each column, and one rhythm lead waveform at the
bottom.
3×4+3
: displays 12-lead ECG waveforms on one page in 4 columns,
with 3 lines in each column, and three rhythm lead waveforms at
the bottom.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...