24 - 2
24.3
Stopping Recordings
Recordings can be stopped manually or automatically.
24.3.1
Stopping Recordings Manually
To manually stop a recording, choose either of the following method:
■
Press the
hardkey again.
■
Select
Clear All Record
in the
Record Setup
menu.
24.3.2
Stopping Recordings Automatically
Recordings stop automatically in the following conditions:
■
The recording is completed.
■
The recorder runs out of paper.
■
The recorder has an alarm condition.
24.4
Recording Related Flags
You can find the following flags on the recording reports:
■
For automatically stopped recordings, there are two columns of asterisks “*” at the end of the report.
■
For manually or abnormally stopped recordings, there is one column of asterisks “*” at the end of the
report.
■
If the parameter data is from external devices connected to the monitor, the parameter label is prefixed
with the plus sign “+”.
24.5
Setting the Recorder
To set the recorder, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Report
column select
Record Setup
.
2.
In the
Record Setup
menu, select the desired waveform for
Waveform 1
,
Waveform 2
and
Waveform 3
in
turn. The recorder can record up to 3 waveforms at a time.
3.
Switch on or off
IBP Overlap
to enable or disable IBP recordings in the overlapping format.
◆
When the
IBP Overlap
is enabled: If two or more waveforms in the selected waveforms for recording
are IBP waveforms, the IBP waveforms will be recorded in the overlapping format.
◆
When the
IBP Overlap
is disabled: IBP waveforms will be recorded normally.
4.
Select
Recording Duration
to set the duration of real-time recording.
5.
Select
Interval
to set the time interval for automatic recording.
6.
Select
Recorder Paper Speed
to set the speed for recording waveforms.
24.6
Enabling Auto Recording on Alarm
To initiate automatic recording via recorder when a parameter alarm is triggered, follow this procedure:
1.
Access the
Alarm
menu for the desired parameter in one of the following ways:
◆
Select the
Alarm Setup
quick key at the bottom of the screen.
◆
Select the numerics area or waveform area of the desired parameter
→
select the
Alarm
tab.
◆
Select the
Parameters Setup
quick key
→
select the desired parameter
→
select the
Alarm
tab.
2.
Switch on
Alarm Outputs
.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...