18 - 10
18.11 Choosing a Time Interval for Peak-Picking
For microstream and mainstream CO
2
modules, you can select a time interval for picking the highest CO
2
as the
EtCO
2
and the lowest as the FiCO
2
.
To set the time interval, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set
Maximum Hold
.
4.
Toggle between
Single Breath
,
10 s
,
20 s
and
30 s
if microstream CO
2
module is configured; toggle
between
Single Breath
,
10 s
and
20 s
if mainstream CO
2
module is configured.
◆
Single Breath
: EtCO
2
and FiCO
2
are calculated for every breath.
◆
10 s
,
20 s
, or
30 s
: EtCO
2
and FiCO
2
are calculated using 10, 20 or 30 seconds of data.
18.12 Changing Barometric Pressure
Both sidestream and microstream CO
2
modules have the function of automatic barometric pressure
compensation (the system automatically measures the barometric pressure to which the patient monitor is
exposed). However, the mainstream CO
2
module does not have such function. For the mainstream CO
2
module,
the default barometric pressure is 760 mmHg. You must modify the barometric pressure based on the actual
situation.
This function is password protected. For more information, see
27.12 The Authorization Setup Settings
.
WARNING
•
Be sure to set the barometric pressure properly before using the mainstream CO
2
module. Improper
settings will result in erroneous CO
2
reading.
18.13 Performing the Leakage Test
When the Sidestream CO
2
module needs maintenance, the monitor will prompt on the CO2 waveform area:
Maintenance Required. Enter CO2 menu.
Then, perform a leakage test by following this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Maintenance
tab.
3.
Perform the leakage test according to the prompt.
18.14 CO
2
Calibration
For sidestream and microstream CO
2
modules, a calibration is needed every year or when the measured values
have a great deviation. For maintream CO2 module, no calibration is needed.
To calibrate the CO
2
module, contact the service personnel.
CAUTION
•
Connect the gas outlet to the scavenging system when calibrating the CO
2
module.
18.15 CO
2
Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service
personnel.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...