18 - 9
18.9
Changing O
2
Settings (for Sidestream CO
2
Module Integrating O
2
)
18.9.1
Changing O
2
Alarm Settings
To change the O
2
alarm settings, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Alarm
tab.
3.
Enter the password if required.
4.
Set alarm properties as desired.
18.9.2
Setting the O
2
Waveform
To set the O
2
waveform, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set
Speed
and
O2 Scale
of the O
2
waveform.
18.10 Setting the Gas Compensation
The presence of interfering gas affects the CO
2
measurement. To get the best possible measuring result, it is
needed to set the gas compensation. The configured concentration of the interfering gas should be in
accordance with its actual proportion.
For the microstream CO
2
module, gas compensations are not required.
WARNING
•
Make sure to use the appropriate compensations. Inappropriate compensations may cause
inaccurate measurement values and result in misdiagnosis.
For the sidestream CO
2
module, follow this procedure to set the gas compensation:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set the compensation according to the actual condition.
For the mainstream CO
2
module, follow this procedure to set the gas compensation:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set the following compensation according to the actual condition.
■
Balance Gas
◆
Select
Room Air
when air predominates in the ventilation gas mixture.
◆
Select
N2O
when N
2
O predominates in the ventilation gas mixture.
◆
Select
He
when He predominates in the ventilation gas mixture.
■
O2 Compensation
◆
Select
Off
when the amount of O
2
is less than 30%.
◆
Select an appropriate setting according to the amount of O
2
in the ventilation gas mixture.
■
AG Compensation
: enters the concentration of anesthetic gas present in the ventilation gas mixture. This
could compensate for the effect of AG on the readings.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...