![Mindray DP-8300 Operator'S Manual Download Page 30](http://html1.mh-extra.com/html/mindray/dp-8300/dp-8300_operators-manual_1800872030.webp)
3. Put the equipotential terminals of the other devices connected to the ultrasonic system to
earth.
WARNING:
1. When you connect another device to this system, you should use
the equipotential wire to connect each of equipotential terminals;
otherwise electric shock may result.
2. Be sure to connect the equipotential wire before inserting the
power plug into the receptacle; be sure to remove the power plug
from the receptacle before disconnecting the equipotential wire;
otherwise electric shock may result.
2.1.3 Connecting/Disconnecting the Transducers
CAUTION:
1. Connect/disconnect the transducer only after the system power is
turned off or the image is frozen (by Freeze key), otherwise failure
may happen.
2. When connecting/disconnecting the transducer, place the
transducer on the corresponding transducer holder and hook the
transducer cable on the cable hanger to avoid accidental falling of
the transducer, which may damage the transducer.
NOTE:
Use the transducer provided by Mindray only. Otherwise may damage the system and
transducer or cause a fire.
Connecting the transducer
WARNING:
Prior to connecting the transducer, the user should ensure that the
transducer, cable and the connector are all in good condition (no rift or
fall-off). Electric shock may happen if using any abnormal transducer.
1. Turn off the ultrasonic system or freeze the image.
2. Check whether the lock of the transducer connector is released. If not, turn the lock
counter-clockwise to release it.
3. Keep the cable upwards and insert the transducer connector straight into the socket.
4. Turn the lock clockwise to lock the transducer connector.
5. Hang the transducer cable to the hanger.
3
4
2-2 Preparing the System
Summary of Contents for DP-8300
Page 1: ...DP 8500 DP 8300 Digital Ultrasonic Diagnostic Imaging System Operator s Manual Basic Volume ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ......
Page 54: ......
Page 58: ......
Page 82: ......
Page 86: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 101: ...P N 046 000308 00 V1 1 ...