7 - 8
BeneVision N1 Patient Monitor Operator’s Manual
7.6.5
Adjusting the Alarm Light Brightness
To adjust the alarm light brightness, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Other
tab.
3.
From the
Alarm Light Brightness
block, set
Primary Display
. If you set alarm light brightness to
Auto
, the
monitor automatically adjusts the alarm light brightness according to the ambient light. The stronger the
ambient light is, the brighter the alarm light is.
For N22/N19, if you are using a secondary display, set the
Secondary Display
too.Restoring the Default Alarm
Settings
To reset all alarm settings to the defaults, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Select
the
Limits
tab
.
3.
On the
Limits
page, select
Defaults
at the left bottom.
7.6.6
Setting the Length of Printing Duration On Alarm
You can define the length of printed waveforms when an alarm is triggered. To do so, follow this procedure:
1.
Access the
Setup
page in either of the following ways:
◆
Select the
Alarm Setup
quick key
→
select the
Setup
tab.
◆
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Select
the
Setup
tab
.
3.
Set
Printing Duration On Alarm
.
7.6.7
Setting the Switch of the CMS and eGateway Disconnection Alarm
You can choose whether to issue an alarm when the monitor is not connected or disconnected from the CMS/
eGateway. To do so, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select
.
2.
Select the
Alarm
tab
→
Other
tab.
3.
Switch on or off
CMS/eGW Disconnected Alarm
. If
CMS/eGW Disconnected Alarm
is switched off, the
“Offline” alarm is not presented when the monitor is not connected or disconnected from the CMS/eGate-
way.
7.6.8
Setting the Priority of the CMS and eGateway Disconnection Alarm
To set the priority of the CMS and eGateway disconnection alarm, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select
.
2.
Select the
Alarm
tab
→
Other
tab.
3.
Set
CMS/eGW Disconnected
.
7.6.9
Setting the Switch of the SpO
2
Desat Alarm Off
You can choose whether switching off the SpO
2
Desat alarm is permissible or not. To do so, follow this
procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .