C - 4
BeneVision N1 Patient Monitor Operator’s Manual
C.1.4
QT Default Settings
C.1.5
Glasgow 12-lead ECG Algorithm Default Settings
C.2
Respiration Default Settings
Item
Default Setting
QTc
Alarm switch
Off
High limit
Adult: 500
Pediatric: 480
Neonate: 460
Priority
Med
Alarm Outputs
Off
∆
QTc
Alarm switch
Off
High limit
60
Priority
Med
Alarm Outputs
Off
QT Analysis
Off
QT Lead
All
Item
Default Setting
Filter
35 Hz
Baseline Drift Removal
On
Tachy
100
Brady
50
Waveform Layout
Standard
Median Complex
Off
Measurements
On
Interpretation
On
Interpretation Summary
On
Item
Default Setting
RR
Alarm switch
On
High limit
Adult: 30
Pediatric: 30
Neonate: 100
Low limit
Adult: 8
Pediatric: 8
Neonate: 30
Priority
Med
Alarm Outputs
Off