22 - 2
BeneVision N1 Patient Monitor Operator’s Manual
22.3
Maintenance and Testing Schedule
Follow the maintenance and testing schedule or local regulations to perform testing and maintenance. Make
sure to clean and disinfect the equipment before taking any tests and maintenance
The following table lists the maintenance and testing schedule:
22.4
Checking Version Information
You may be asked for information on monitor and module version.
To view system software version information, select the
Main Menu
quick key
→
from the
System
column select
Version
.
You can also view more version information by following this procedure
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select
.
2.
Select the
Version
tab.
You can check system software version, module hardware and software version, and firmware version.
22.5
Testing Methods and Procedures
Except the following maintenance tasks, all other test and maintenance tasks should be performed by Mindray-
qualified service personnel only.
■
Regular check, including visual inspection and power-on test
■
Printer tests
■
Battery check
If your monitor needs a safety test and performance test, contact the service personnel.
Test/Maintenance Item
Recommended Frequency
Performance Tests
Visual inspection
Every day, before first use.
Measurement module performance test and
calibration
1. If you suspect that the measurement values are incorrect.
2. Follow any repairs or replacement of relevant module.
3. Once a year for CO
2
tests.
4. Once every two years for other parameter module performance
tests.
Analog output test
If you suspect that the analog output function does not work
properly.
Defibrillation synchronization test
If you suspect that the defibrillation synchronization function does
not work properly.
Electrical Safety Tests
Electrical safety tests
Once every two years.
Other Tests
Power-on test
Before use.
Network printer tests
1. When first installed.
2.Followany repair or replacement of the printer.
Battery check
Functionality test
1. When first installed.
2. When battery is replaced.
Performance test