Maintenance
25-5
25.7 Calibrating AG
Calibrate the AG module every year or when the measured value has a great deviation.
Tools required:
Gas bottle, with a certain standard gas or mixture gas. Gas concentration should meet
the following requirements: AA>1.5%, CO
2
>1.5%, N
2
O>40%, O
2
>40%, of which AA
represents an anesthetic agent.
T-shape connector
Tubing
Reservoir bag
Follow this procedure to perform a calibration:
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance
>>]
→
enter the
required password
→
[
Calibrate AG >>
].
2. Check the airway and make sure that there are no occlusions or leaks.
Vent the tubing to the air and check if the [
Current FlowRate
] and [
Set FlowRate
]
are approximately the same. If the deviation is great, it indicates that there is an
occlusion in the tubing. Check the tubing for an occlusion.
Block the gas inlet of the tubing. The [
Current FlowRate
] shall fall rapidly and
the system prompt that the tubing is blocked. Otherwise, it indicates that there are
leakages in the tubing. Check the tubing for leakages.
3. Connect the gas bottle, reservoir bag and the tubing using a T-shape connector as shown
in the figure above. Check the airway and make sure there are no leaks.
4. Vent the tubing to a certain standard gas or mixture gas opening the gas valve.
5. In
the
[
Calibrate AG
] menu, the measured gas concentration and flowrate are
displayed.
If the difference between the measured gas concentration and the actual one is very
small, a calibration is not needed.
AG module
Reservoir bag
Gas bottle
Gas valve
Tubing
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...