Maintenance
25-4
25.6 Calibrating CO
2
For sidestream and microstream CO
2
modules, a calibration is needed every year or when the
measured values have a great deviation. For maintream CO
2
module, no calibration is
needed.
Tools required:
Gas bottle, with 3%, 4%, 5%, 6% or 7% of CO
2
.
T-shape connector
Tubing
Follow this procedure to perform a calibration:
1. Make sure that the CO
2
module has been warmed up or started up.
2. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance
>>]
→
enter the
required password
→
[
Maintain CO2 >>
].
3. In
the
[
Maintain CO2
] menu, select [
Zero
].
4. Connect the gas bottle with the tubing using a T-shape connector as shown below.
Check the airway and make sure there are no leaks.
5. Vent the tubing to the CO
2
opening the gas valve.
6. In
the
[
Calibrate CO2
] menu, enter the vented CO
2
concentration in the [
CO2
] field.
7. In
the
[
Maintain
CO2
] menu, the measured CO
2
concentration, barometric pressure,
sensor temperature and current flowrate are displayed. After the measured CO
2
concentration becomes stable, select [
Calibrate CO2
] to calibrate the CO
2
module.
8. If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. If the calibration failed, the message [
Calibration Failed!
] is displayed.
Perform another calibration.
9. Select to exit the current menu.
Monitor
Gas bottle
Open to the air
Gas valve
Tubing
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...