Monitoring BIS
16-4
16.4 Setting up the BIS Measurement
1. Connect the BISx model to the BIS module.
2. Use the attachment clip to secure the BISx model near, but not above the level of the
patient’s head.
3. Connect the BISx model to the patient cable.
4. Attach the BIS sensor to the patient following the instructions supplied with sensor.
NOTE
z
Make sure the patient’s skin is dry. A wet sensor or a salt bridge could result in
erroneous BIS and impedance values.
5. Connect the BIS sensor to the patient interface cable. As soon as a valid sensor is
detected, the impedances of all electrodes are measured automatically and the
impedance value for each electrode is displayed in the sensor check window.
CAUTION
z
Do not attach the BISx model to the patient’s skin for long time. Otherwise, the
BISx heats while on the patient and may cause discomfort.
Patient cable
BIS module
BISx model
BIS sensor
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...