7 - 6
Defibrillator/Monitor Operator’s Manual
5.
Place the conductive surface of paddle electrodes against the patient's right atrium and left ventricle, as
shown in the figure below:
6.
Charge the defibrillator by pressing the Charge button on the front panel.
7.
Make sure no one is touching the patient or anything connected to the patient.
8.
Press the Shock button on the front panel.
NOTE
•
When internal paddles are used for defibrillation, the energy selection is automatically limited to 50
joules because of possible cardiac damage from higher energies.
•
Sterilize the internal paddles before each use. Otherwise, severe infection may result.
•
Clean the internal paddles after each use.
7.5
Synchronized Cardioversion
Synchronized cardioversion allows you to synchronize delivery of the defibrillator shock with the R-wave of the
ECG. You may choose to perform synchronized cardioversion through electrode pads, external and internal
paddles.
To use synchronized cardioversion, press the [
Enter Sync
] soft key in the asynchronous defibrillation mode.
Then
[SYNC]
appears in the manual Defibrillation information area and a marker appears above each R-wave,
see the figure below:
You can monitor ECG through electrode pads, or external paddles, or electrodes connected to a 3- or 5-lead ECG
cable. Shock is delivered through either electrode pads or external paddles. For synchronized cardioversion, we
recommend to acquire patient’s ECG through ECG lead set.
SYNC marker
R-wave marker
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...