F - 2
Defibrillator/Monitor Operator’s Manual
F.3
Device Labeling
Check the labels provided by the manufacturer or the healthcare facilities are present and legible.
■
Main unit label
■
Integrated warning labels
F.4
Protective Earth Resistance
1.
Plug the probes of the analyzer into the device’s protective earth terminal and protective earth terminal of
the AC power cord.
2.
Test the earth resistance with a current of 25 A.
3.
Verify the resistance is less than limits.
LIMITS
For all countries, R = 0.2 Ω Maximum
F.5
Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage tests.
The following outlet conditions apply when performing the Earth Leakage test:
■
normal polarity (Normal Condition)
■
reverse polarity (Normal Condition)
■
normal polarity with open neutral (Single Fault Condition)
■
reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
◆
300 μA in Normal Condition
◆
1000 μA in Single Fault Condition
For IEC60601-1,
◆
500 μA in Normal Condition
◆
1000 μA in Single Fault Condition
F.6
Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All measurements have
a true RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
■
normal polarity (Normal Condition)
■
reverse polarity (Normal Condition)
■
normal polarity with open neutral (Single Fault Condition)
■
reverse polarity with open neutral (Single Fault Condition)
■
normal polarity with open earth (Single Fault Condition)
■
reverse polarity with open earth (Single Fault Condition)
LIMITS
For CF
applied parts
◆
10μA in Normal Condition
◆
50μA in Single Fault Condition
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...