Defibrillator/Monitor Operator’s Manual
A - 11
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2)in
terms of mean error and stardard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
stardard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
A.5.7
CO
2
Specifications
Auto mode repetition intervals
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time
5 min
Static pressure measurement range
0mmHg to 300mmHg
Static pressure measurement accuracy
±3mmHg
Maximum measurement time
Adult, Pediatric:
Neonate:
180s
90s
Initial cuff inflation pressure range
Adult:
Pediatric:
Neonate:
80 to 280 mmHg
80 to 210 mmHg
60 to 140 mmHg
Default Initial cuff inflation pressure
Adult:
Pediatric:
Neonate:
160 mmHg
140 mmHg
90 mmHg
Measurement range
Adult
Pediatric
Neonate
Systolic
mmHg
25 to 290
25 to 240
25 to 140
Diastolic
mmHg
10 to 250
10 to 200
10 to 115
Mean
mmHg
15 to 260
15 to 215
15 to 125
Software overpressure protection
Adult:
297±3 mmHg
Pediatric:
297±3 mmHg
Neonate:
147±3 mmHg
Measurement accuracy*
Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution
1 mmHg
Measurement range
0 to 150 mmHg
Accuracy*
Full accuracy mode:
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
100 to 150 mmHg:
±2 mmHg
±5% of reading
±10% of reading
±(3 mmHg+8% of reading)
ISO accuracy mode: Add ±2mmHg to the full accuracy mode
Sart-up time
20 s (typical), 90 s (maximum)
Accuracy drift
Meets the requirement for measurement accuracy within 6 hours.
Resolution
1mmHg
Sample flowrate
Connecting the Oridion sampling line: 50 ml/min
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...