Defibrillator/Monitor Operator’s Manual
A - 7
A.5.2
ECG Specifications (from Defibrillation Electrodes)
Time to alarm for tachycardia
Meets the requirements in Clause 201.7.9.2.9.101 b) 6) of IEC 60601-2-27.
Waveform
4ah - range:
4a - range:
4ad - range:
4bh - range:
4b - range:
4bd - range:
<11 s
<11 s
<11 s
<11 s
<11 s
<11 s
Arrhythmia Analysis Classifications
Asystole, V-Fib/V-Tach, V-Tach, Vent. Brady, Extreme Tachy, Extreme Brady,
PVCs/min, PVC, Couplet, VT>2,Bigeminy, Trigeminy, R ON T, Brady, Tachy,
Missed Beat, PNP, PNC, Vent Rhythm, Multif. PVCs, Nonsus. Vtac, Pause, Irr.
Rhythm, A-Fib
Tall T-wave rejection capability
When the test is performed based on Clause 201.12.1.101.17 of IEC 60601-2-
27, the heart rate meter will reject all 100 ms QRS complexes with less than
1.2 mV of amplitude, and T waves with T-wave interval of 180 ms and those
with Q-T interval of 350 ms.
Lead-off detection current
Measuring electrode:
Drive electrode:
≤0.1 μA
≤1 μA
Baseline recovery time
<2.5 s (after defibrillation)
Response to irregular rhythm
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4) of IEC
60601-2-27, the heart rate after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy (3a): 80±1 bpm
Slow alternating ventricular bigeminy (3b): 60±1 bpm
Rapid alternating ventricular bigeminy (3c): 120±1 bpm
Bidirectional systoles (3d): 90±2 bpm
Patient connection
paddles or multifunction electrode pads
ECG inputs
pads/paddles
Gain
2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1), 20 mm/mV (×2),
40mm/mV (×4), Auto. Error less than
±
5%
Sweep speed
6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s. Error no more than
±
5%
Bandwidth (-3dB)
Therapy mode:
1 to 20 Hz
Common mode rejection
Therapy mode:
>105 dB
Notch filter
50/60Hz
In Therapy mode:
notch filter turns on automatically
ECG signal range
±8mV (peak-to-peak value)
Calibration signal
1mV (peak-to-peak value) ±5%
Differential input impedance
≥5 MΩ
Electrode offset potential tolerance
±1V
Defibrillation protection
Enduring 5000V (360 J) charge without data loss or corruption
Baseline recovery time: <2.5 s (after defibrillation)
Polarization recovery time: <10 s
Defibrillation energy absorption: ≤10% (100Ω load)
ESU protection
Cut mode:
Coagulate mode:
Recovery time:
300 W
100 W
≤10 s
In compliance with the requirements in clause 202.6.2.101 of IEC 60601-2-27
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...