21 - 2
Defibrillator/Monitor Operator’s Manual
21.4
Exporting Configurations
1.
Connect the USB flash memory to the equipment’s USB connector.
2.
Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
3.
Press the Main Menu button on the front panel.
4.
Enter the Configuration Main menu by either of the following ways:
◆
Select [
Others >>
]
→
[
Configuration >>
]
→
[
View Config
].
◆
Select [
Others >>
]
→
[
Configuration >>
]
→
enter the required password
5.
Select [
Config. Export
].
21.5
Importing Configurations
It is not necessary to configure each equipment separately when installing several equipments with identical
configurations. A USB flash memory can by used to import configurations from one equipment to another.
1.
Prepare a USB flash memory with desired configurations.
2.
Connect the USB flash memory to the target equipment’s USB connector.
3.
Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
4.
Press the Main Menu button on the front panel.
5.
Enter the Configuration Main menu by either of the following ways:
◆
Select [
Others >>
]
→
[
Configuration >>
]
→
[
View Config
].
◆
Select [
Others >>
]
→
[
Configuration >>
]
→
enter the required password
6.
Select [
Config. Import
]
21.6
Changing Configurations
You can change configurations either by changing settings in each configuration setup menu, or restoring to
factory defaults.
1.
Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
2.
Press the Main Menu button on the front panel, and then select [
Others >>
]
→
[
Configuration >>
]
→
enter the required password.
◆
Selecting [
Factory Default
] can restore all the current settings to factory defaults.
◆
Select the desired setup menu, and change the settings.
◆
Selecting [
Record
] can record all the current settings.
21.6.1
General Setup Menu
Menu Item
Options/Range
Default
Description
Device Name
20 characters
/
Inputs the relevant information.
The characters are included in
the keyboard. Restoring factory
default configurations does not
change these settings.
If the equipment is connected to
the CMS, [
Bed No.
] and
[
Department
]can also be
changed on the CMS.
Institution
20 characters
/
Department
20 characters
/
Bed No.
20 characters
/
Patient Cat.
Adu, Ped, Neo
Adu
Sets the default patient category.
Patient Info Display
Bed No., Patient Name
Bed No.
Sets the patient information type
displayed in the upper left corner
of the main screen.
Height Unit
cm, inch
cm
Sets the default measurement
unit for the patient.
Weight Unit
kg, lb
kg
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...