11-1
11
Maintenance
WARNING
To avoid electric shock, stop using the device if you find its housing has
signs of broken. Contact the service personnel for help in that case.
The hospital or medical facility using this infusion pump must set up a
comprehensive maintenance plan. Failure to do so may result in equipment
failure or other unexpected consequences, and may even jeopardize
personal safety.
All safety inspections or maintenance work involving the disassembly of
the device must be conducted by professional maintenance personnel.
Actions by unqualified persons may result in device failure and may even
jeopardize personal safety.
Please contact the company immediately if you encounter problems with
the device (such as the warning label off).
The device and accessories shall not be served or maintained while in use
with a patient.
11.1 Inspection
The infusion pump must be given a thorough inspection before use, after 6-12
months of continuous use, and after maintenance or upgrades, to ensure that it is
operated and functioned normally.
The inspection criteria are:
The environment and power supply meet requirements.
The equipment and accessories have no mechanical damage.
The power cord is not damaged and has sufficient electrical insulation.
Accessories used with the pump are correct.
The alarm system functions correctly.
Battery performance.
Self-checking and pump functions are normal.
If there are any forms of damage or abnormal circumstances, do not use the infusion
pump and contact the company immediately.
Summary of Contents for BeneFusion VP5
Page 1: ...BeneFusion VP5 BeneFusion VP5 Ex BeneFusion VP5 Neo Infusion Pump Operator s Manual ...
Page 2: ......
Page 83: ...A 12 Unit of Rate ml h ...
Page 102: ...F 1 F Declaration of Conformity P N 046 013877 00 1 0 ...
Page 103: ......
Page 104: ......