6-6
6.10.3 Infusion Pump with PVC Specified Infusion Sets
1. Select [
Main Menu
]→[
User Maintenance
]→Input User Maintenance
Password→[
Infusion Set
].
2. Select in [
Infusion Set
] according to actual needs.
Note: Please ensure that at least one “Infusion Set” to be selected.
Brand and model of PVC specified infusion sets:
No.
Type
Brand
Model
1
Regular
SK
SK ZPQ
2
Regular
SK
SK 150mL
3
Light-sensitive
SK
SK JMB
6.11 Set the Sensitivity of Empty Bottle Alarm
1. Select [
Main Menu
]→[
General Options
]→[
Empty Sensitivity
].
2.
Select [
Empty Sensitivity
]:
High
and
Low
. [
High
] indicates high sensitivity of
[
Empty
] alarm detection, [
Empty
] alarm will be triggered in shorter time; [
Low
]
indicates low sensitivity of [
Empty
] alarm detection, [
Empty
] alarm will be triggered
in longer time. It can be selected according to actual needs
.
Note: [
Empty Sensitivity
] option is not visible when the switch of [
Drop sensor
] is
off.