4-8
CAUTION
Please confirm that the current selected brand is the same as the brand
actually used, or its accuracy cannot be guaranteed.
4.2.6 Parameter Memory Function
In clinical treatments, the medical staffs need to initiate the infusion as soon as
possible during emergency situations, infuse the liquid drug into the patient's body in
the shortest time possible, and set the detailed parameters later during infusion.
1. Select
[
Main Menu
]→[
User Maintenance
]→Input User Maintenance
Password→[
Para.
Memory
].
2. Select [
Para. Memory
]→[
On
]. If [
Off
] is selected, the following steps cannot be
conducted.
3. After selecting the infusion set brand, the previous infusion screen will appear, the
previous therapy parameters will be loaded, the users can use the previous
treatment parameters.
4.2.7 Select Infusion Mode
Press
to enter [
Main Menu
]
→[
Select Mode
]. On this interface, users can press
and
to select infusion mode. Please refer to
Chapter 5 Infusion
Mode
for detailed introduction of each infusion mode.
4.2.8 Purge Air
During infusion, the user should prevent air bubbles from entering the blood with the
liquid drug, which may form an aeroembolism and put patients in serious danger.
Therefore, air bubbles in the infusion tube should be eliminated before the infusion.
Under non-running status of any infusion mode, press
to enter [
Purge Air
]
setting screen, set [
Purge Rate
], then hold down
to purge at the set rate,
release
after the air bubbles are purged.
WARNING
Please disconnect the pump from patient before purge. Otherwise, the
patient will be in serious danger
!