4-2
4.2 Operational Procedures
4.2.1 Turn on the Pump
To turn on the device, follow this procedure:
1. Perform a safety inspection referring to
11.1 Inspection
before turning on the
pump.
2. Press
, the system will initiate the self-test and the screen will display the
[
System Self-test
] interface:
System will give out a sound "di"- indicating the self-test of the loudspeaker to be
successful.
The color of the alarm indicator light will change from red to yellow, turn on and
off orderly- indicating the self-test of the alarm light to be successful.
System will give out a sound "didi"- indicating the self-test of the buzzer to be
successful.
3. Enter the operation interface after successfully completing the system self-test,
and now you can operate the system through the key board.
WARNING
Please monitor the self-test process to make sure that the speaker, the
alarm light, and the buzzer are all self-tested successfully. Otherwise,
please contact the company and do not operate the pump until
maintenance is performed.
Please contact the company if the infusion pump is damaged or cannot
operate properly, and it cannot be used for patient infusion.
4.2.2 Load the Infusion Set
4.2.2.1 Infusion Pump with Conventional Infusion Sets
System will inspect whether the infusion set is loaded after completing the self-test:
If infusion set is not loaded, enter the infusion set [
Loading Guide
] interface; If
the infusion set is not required to load, press
to skip the step.
If infusion set is loaded incorrectly and door is closed, enter the [
Loading Error
]
interface.
If infusion set is correctly loaded and door is closed, and [
User Maintenance
]→
[
Brand Selection
]
→[
On
], enter the [
Set Selection
] interface; if the switch of
[
Brand Selection
] is [
Off
], then skip the [
Set Selection
] interface.