3-2
WARNING
The packaging materials must be kept out of the reach of children. They
must be disposed of in compliance with local laws and regulations or
the hospital policy on waste management.
The equipment may be contaminated by microbes during storage,
transport and use. Please ensure that the package is undamaged before
using, do not use if there is any damage.
NOTE
Keep the packing case and packaging materials for future transportation
or storage.
3.1.2 Operating Conditions
The operating environment of this infusion pump must meet the requirements in
A.1.2
Operating Environment
.
The operating environment should also be appropriately protected from noise,
vibration, dust, and corrosive, inflammable or explosive substances. If installed inside
the equipment case, a sufficient space before and after the equipment case should
be ensured to facilitate operation, maintenance and repairing work. There should be a
2" (5 cm) gap around the infusion pump to ensure that air can circulate freely for a
better cooling effect.
When the pump is transferred from one place to another, differences in temperature
and humidity can cause condensation to form inside the pump. If this occurs, do not
switch the pump to the "ON" state until the condensation has gone.
WARNING
Please use only when the operating environment meets the
requirements specified above. Otherwise, the pump's performance will
not match the technical specifications in A Product Specifications. Device
failure and other unexpected consequences may also result.