5-10
9. As shown in the figure: Remove the display.
5.2.5 Disassemling the Pump Body Assembly (Note: You
should wear an anti-static wristband throughout)
1. As shown in the figure: Pull out the hollow shaft sleeve by hand.
2. As shown in the figure: Loosen the two M3X8 cross small pan head combination
screws with electric screwdriver and remove the motor from the pump body
assembly.
Hollow shaft sleeve