5-7
5.2.4 Disassembling Front Cover Assembly (Note: You should
wear an anti-static wristband throughout)
1. As shown in the figure: Remove the two M3x8 Phillips pan head double washer
screws that locks the core control board PCBA with a Phillips screwdriver.
2. As shown in the figure: Unplug the speaker cable and remove the core control board
PCBA.
3. As shown in the figure: use the Phillips screwdriver to remove the six M3x8 Phillips
pan head double washer screws that fix the keypad board PCBA, and take out of the
keypad board PCBA. Handle gently, protect the display during removal, and prevent
the display adhesive from being disengaged.
Summary of Contents for BeneFusion SP3
Page 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Page 2: ......
Page 12: ...1 6 FOR YOUR NOTES ...
Page 24: ...2 12 FOR YOUR NOTES ...
Page 62: ...5 16 FOR YOUR NOTES ...
Page 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Page 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Page 78: ...7 8 FOR YOUR NOTES ...
Page 79: ......
Page 80: ...P N 046 013115 00 3 0 ...