5 - 2
5.2
Relay Infusion
Multiple pumps can be combined to infuse at a preset sequence when used with the
Dock. Pumps in a single Dock or pumps in the cascade Docks are all available for relay
infusion.
NOTE
•
Relay infusion is available for Rate Mode, Micro-infusion Mode, Time Mode,
and Dose Mode.
•
When a relay infusion is set up, the sequence of the current pump in the rely
is displayed in the system information area. For example, symbol
indicates that the current pump is the second in a circular relay.
•
You cannot change the sequence of the pumps when the relay infusion is set
up.
5.2.1
Setting up a Relay Infusion
To set up a relay infusion, follow this procedure:
1.
Connect the pump to the Dock.
2.
Swipe the touchscreen from top down, and select
Relay
.
3.
Select one of the following options:
◆
Customized Relay
: The relay infusion runs in a preset order, and completes
when the last relay pump finishes the infusion.
◆
Circular Relay
: The relay infusion runs in a preset order, and the first pump
continues to infuse when the last relay pump completes the infusion.
4.
From the desired pumps, select
Yes
in the dialog.
5.
Select the sequence of the relay pumps.
6.
Select
Confirm
from the initial pump to complete the setting.
7.
Set the infusion parameters respectively for the relay pumps.
8.
Select
from the first pump to start the relay infusion.
5.2.2
Canceling the Relay
To cancel a relay infusion, swipe the touchscreen from top down
→
select
Relay
→
select
Cancel
.
■
For the circular relay, canceling at the current pump removes all pumps from the
relay infusion.
■
For the customized relay, canceling at the current pump removes all pumps from
the relay infusion.
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...