4 - 9
•
To prevent free flow, make sure that the clamp has fully occluded the
extension set before unloading a syringe.
4.11
Viewing the Infused Volume
The
Volume
dialog allows you to review the infused volume of up to 24 hours. You can
also view the infused volume of the configured time interval and time length.
Choose either of the following ways to enter the
Volume
dialog:
■
Swipe the touchscreen from top down
→
select
Volume
.
■
Select
Volume
from the
Pause
screen.
NOTE
•
The infusion volume cannot be cleared when an infusion is running.
(1)
(2)
(3)
(4)
(5)
(1)
Past 24h Total
: view the total infused volume in the past 24 hours. The display range
is 0 ml to 99999.99 ml.
Select
Clear
to clear the infused volume.
(2)
View the total infused volume in the configured time period.
Configure the time period before viewing the total infused volume in the configured
time period.
(3)
View the recent total infused volume.
Configure the time before viewing the total infused volume within the configured
time.
(4)
Timing Volume
: view the total infused volume of the configured timing interval.
Configure the
Timing Interval
before viewing the total infused volume of each
interval.
(5)
History Rate
: view the history rate.
Summary of Contents for BeneFusion eSP Vet
Page 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Page 2: ......
Page 12: ...6 This page intentionally left blank ...
Page 26: ...2 10 This page intentionally left blank ...
Page 44: ...5 4 This page intentionally left blank ...
Page 74: ...11 8 This page intentionally left blank ...
Page 82: ...12 8 This page intentionally left blank ...
Page 100: ...A 10 This page intentionally left blank ...
Page 108: ...B 8 This page intentionally left blank ...
Page 111: ...D 1 D Declaration of Conformity ...
Page 112: ...D 2 This page intentionally left blank ...
Page 113: ......
Page 114: ...P N 046 021887 00 1 0 ...