3 - 4
Operator’s Manual
3 System Preparation
Perform the following procedure:
1.
Click
in the bottom-right corner to bring up the Session Manage dialog box where you can
see the current user’s information.
2.
If you want to modify the current password, click [Change Password] to bring up the Change
Password dialog box.
3.
Enter both the previous and new passwords, and confirm the new password in the dialog box.
4.
Click [OK] to exit.
To lock the system
Perform the following procedure:
1.
Click the
in the bottom-right corner of the screen to bring up the dialog box.
2.
Select [Lock Machine] and the system is locked.
You must log on before using the system.
3.3.3 Check the system after it is powered on
Check after the system is powered on:
•
There are no unusual sounds or smells indicating possible overheating.
•
There are no persistently displayed system error messages.
•
There is no evident excessive noise, or discontinuous, absent or black items in the B mode
image.
•
Check whether there is abnormal heat on the surface of the probe during an ultrasound
procedure. If you use a probe which is giving off excessive heat, it may burn the patient.
•
The control panel keys and knobs are fully functional.
•
The date and time are displayed correctly.
WARNING
•
If you use a probe giving off excessive heat, it may burn the patient.
•
If you find anything not functioning properly, this may indicate that the
system is defective. In this case, shut down the system immediately and
contact Mindray Customer Service Department or sales representative.
NOTE:
When you start the system or switch between transducers, you will hear clicking sounds – this is
expected behavior.
3.3.4 Power the System Off
You must follow the correct procedures to power the system off. Also, after you upgrade the
software or when the system is down, you need to power off and restart it.
If you will not use the system for a long period of time, you shall:
•
Disconnect the power adapter.
•
Disconnect the mains power.
•
Turn off powers of all peripherals connected to the system.
Summary of Contents for Ana
Page 2: ......
Page 50: ...This page intentionally left blank...
Page 60: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 116: ...This page intentionally left blank...
Page 166: ...This page intentionally left blank...
Page 176: ...This page intentionally left blank...
Page 194: ...This page intentionally left blank...
Page 220: ...This page intentionally left blank...
Page 288: ...This page intentionally left blank...
Page 304: ...This page intentionally left blank...
Page 308: ...This page intentionally left blank...
Page 316: ...This page intentionally left blank...
Page 337: ......
Page 338: ...P N 046 018835 00 2 0...