Accutorr V Operating Instructions
ix
Notes
Introduction
Notes
NOTE:
The Accutorr V should be operated only by trained and
qualified personnel.
NOTE:
Use disposable and single use accessories only once.
NOTE:
Place the equipment in a location where the screen can
easily be seen and the operating controls can easily be
accessed.
NOTE:
In certain situations in which perfusion and signal strength
are low, such as in patients with thick or pigmented skin,
inaccurately low SpO
2
readings will result. Verification of
oxygenation should be made, especially in preterm infants
and patients with chronic lung disease, before instituting
any therapy or intervention.
NOTE:
The instructions in this manual are based on the maximum
configuration.
NOTE:
The optional Temperature module kit must be installed only
by trained personnel, and proper ESD prevention methods
must be followed.
NOTE:
Only devices specified by Mindray DS USA, Inc.
/Shenzhen Mindray
Bio-Medical Electronics Co., Ltd shall be
connected the RS-232 port
.
NOTE:
When the RS-232 connector is used for DIAP, barcode power
must be set to OFF.
NOTE:
Disconnect the Accutorr V from the mains to isolate it from
the mains power during an emergency.
Safety Designations
Safety designations per IEC 60601-1 Standard:
Type of protection against electric shock
Class 1 with internal electric power source.
Where the integrity of the external protective
earth (ground) in the installation or its
conductors is in doubt, the equipment shall be
operated from its internal electric power
source.
Degree of protection against electric shock
Monitor - Type B applied part.
NIBP - Type BF defibrillation protected
applied part.
SpO
2
- Type BF protected applied part.
Temp - Type BF protected applied part.
Supply Connection
100 – 240 VAC
50/60 Hz
0.85 – 0.5 A