36
MW170 Bench Meter
Interval Logging
1. From the Setup mode, set Log Type to INTERVAL (default) and desired time interval.
2. From the measurement screen press LOG/CLEAR.
LCD displays “PLEASE WAIT”. The LOG ### LOT ### screen displays on third LCD line
the measurement log number (bottom left) and interval logging session lot number
(bottom right).
3. Press RANGE/
during logging to display the number of available records
("FREE" ###). Press RANGE/
again to return to return to active logging screen.
4. Press LOG/CLEAR again (or ESC) to end current interval logging session.
LCD displays “LOG STOPPED”. Meter returns to measurement screen.
Interval Logging Warnings
“OUT OF SPEC”
Sensor failure is detected. Loggings stops.
“MAX LOTS”
Maximum number of lots reached (100). Cannot create new lots.
“LOG FULL”
Log space is full (1000 logs limit was reached). Loggings stops.
10.2. DATA MANAGEMENT
y
A lot contains 1 to 600 log records (saved measurement data points)
y
Maximum number of lots that can be stored is 100, excluding Manual and Stability
y
Maximum number of log records that can be stored is 1000, across all lots
y
Manual and Stability logs can store up to 200 records (each)
y
Interval logging sessions (across all 100 lots) can store up to 1000 records.
When a logging session exceeds 600 records a new lot will be created.