OPERATING 

• Remove the 
protection cap.  
• Turn the CD97 on 
with the ON/OFF 
switch located on the 
top. 
• Immerse into solution 
up to the maximum 
immersion level without 
touching the bottom of 
the beaker. 
• Stir gently and wait 
until the display 
stabilizes. 

CD97

 

compensates for the 
temperature variance 
automatically. 
• Read the value on the 
display.

         

 

CALIBRATION 

• The calibration 
procedure is very 
simple and 
fast. 
• Immerse 
the tester up 
to the 
maximum 
level into the 
appropriate 
optional 
calibration 
solution 
M10080 
corresponding to 800 
ppm. 
• Allow the displayed 
value to stabilize. 

• Pull out the battery 
compartment and 
locate the calibration 
trimmer on the top of 
the internal circuitry. 
• By turning the 
calibration trimmer with 
a small screwdriver, 
adjust the reading to 
match the solution 
value of 800 ppm 
(normally to the value 
at 25°C). Replace the 
battery compartment in 
the original position. 

 

• Calibration is now 
complete.

  

 

BATTERY 
REPLACEMENT 

When the 

CD97

 

cannot be 
switched on or 
the display 
fades, pull out 
the battery 
compartment 
and replace all 
three 1.5V batteries, 
paying attention to their 
polarity. Batteries 
should be replaced in a 
non-hazardous area 
using the battery types 
specified in this 
instruction manual. 

RECOMMENDATIONS FOR 
USERS

Before using this product, make 
sure that it is entirely suitable for 
the environment in which it is used. 
Operation of this instrument in 
residential area could cause 
unacceptable interferences to radio 
and TV equipments, requiring the 
operator to take all necessary steps 
to correct interferences. 
The glass bulb at the end of the pH 
electrode is sensitive to 
electrostatic discharge. Avoid 
touching this glass bulbat all times. 
During operation of instrument, 
ESD wrist straps should be warn to 
avoide possible damage  to the pH 
electrode  by electrostatic 
discharge. Any variation introduced 
by the user to the supplied 
equipment may degrade the 
instrument’s EMC performance. To 
avoid electrical shock, do not use 
this instrument when voltages at the 
measurement surface exceed 24 
VAC or 60 VDC. To avoid damages 
or burns, do not perform any 
measurement in microwave ovens. 
 

SPECIFICATIONS 
Range

   

0 to 999 ppm 

Resolution

 1 

ppm 

Accuracy

     

±10 ppm 

Temperature

 automatic, 

Compensation

 from 

 

  5 

to 

50° 

 

Environment

 

0 to 50°C 

Battery Type

 4x1.5V 

alkaline 

Battery life

 150 

hours 

Dimensions

 

150x30x24 mm 

Weight

  

85 g

 

 