Sterylis VS Manual
| 24
If the room exceeds the 10 ppm ozone concentration threshold, the unit will display an error and proceed to ozone
destruction until the safe concentration is reached and the safe ozone concentration light is lit.
6.3.4.5 Emergency situations during the sterilization process
The following emergencies may occur during operation in sterilization mode:
Emergency situation
Behavior of the device after an emergency
situation
The sterilization process started before the time
required to heat up the ozone sensor (default 30 min)
Display error P005, return to standby mode
During the ozoning process, an ozone concentration
of 10 ppm was exceeded
Display of error P003, interruption of the process
and transition to the ozone destruction phase until
the safe concentration is reached
A concentration of 0.5 ppm will not be reached within
120 minutes from the start of the sterilization
process
Display of error P004, interruption of the
sterilization process and transition to the ozone
destruction phase until the safe concentration is
reached
A concentration of 0.1 ppm will not be reached within
15 minutes from the start of the sterilization process
Display of error E501, interruption of the
sterilization process and transition to the ozone
destruction
Ozone sensor malfunction
Display of error E500 or E501, interruption of the
sterilization process and transition to the ozone
destruction phase
It is also possible to short circuit the relay output and display error E300. This is a critical error!
If this occurs, the power supply of the device should be disconnected as safely as possible,
preferably by means of an overcurrent switch in the switchgear or, if necessary, by unplugging
the plug of the power cord from the electrical outlet in accordance with the safety rules
described in these operating instructions! After disconnecting the power supply, contact the
service department.
6.4 Sleep mode
The room sterilizer STERYLIS VS goes into so-called sleep mode when it is connected to the power supply and the main
switch is in the OFF position. This is a mode in which the device looks off - the user panel is completely extinguished, no
errors are signalled and no fans, UV lamps and ozone generators are running. However, the concentration of ozone is
monitored. Possible exceeding of the safe ozone concentration will be indicated by a warning beacon light and indicated
on the display in the CONCENTRATION section.
6.5 Front panel locking mode
The STERYLIS VS room sterilizer is equipped with the possibility of activating a front panel lock. It is intended to block the
device in such a way that undesirable persons cannot operate it, including in particular the possibility of accidentally
starting a sterilization mode. In lock mode, all buttons on the front panel (except those used to exit the mode) are no
longer active.
Lock mode can only be activated when the unit is in standby, silent disinfection, standard disinfection or filtration mode.
Lock mode features:
After entering and leaving the sleep mode, the lock mode remains active
The mode is active without any time limit
When the device is restarted, the lock mode is always inactive
In the lock mode, the display in the O
3
STERILIZATION section shows "L0CH" (scrolling with possible error codes