background image

 

 

 

Sterylis VS Manual                                                                                                                                                                                       

| 24

 

 

If  the  room  exceeds  the  10  ppm  ozone  concentration  threshold,  the  unit  will  display  an  error  and  proceed  to  ozone 
destruction until the safe concentration is reached and the safe ozone concentration light is lit. 

6.3.4.5 Emergency situations during the sterilization process

 

The following emergencies may occur during operation in sterilization mode: 
 

Emergency situation 

Behavior  of  the  device  after  an  emergency 
situation 

The  sterilization  process  started  before  the  time 
required to heat up the ozone sensor (default 30 min) 

Display error P005, return to standby mode 

During the ozoning process, an ozone concentration 
of 10 ppm was exceeded 

Display  of  error  P003,  interruption  of  the  process 
and transition to the ozone destruction phase until 
the safe concentration is reached 

A concentration of 0.5 ppm will not be reached within 
120  minutes  from  the  start  of  the  sterilization 
process 

Display  of  error  P004,  interruption  of  the 
sterilization  process  and  transition  to  the  ozone 
destruction  phase  until  the  safe  concentration  is 
reached 

A concentration of 0.1 ppm will not be reached within 
15 minutes from the start of the sterilization process 

Display  of  error  E501,  interruption  of  the 
sterilization  process  and  transition  to  the  ozone 
destruction  

Ozone sensor malfunction 

Display of error E500 or E501, interruption of the 
sterilization  process  and  transition  to  the  ozone 
destruction phase 

It is also possible to short circuit the relay output and display error E300. This is a critical error! 
If  this  occurs,  the  power  supply  of  the  device  should  be  disconnected  as  safely  as  possible, 
preferably by means of an overcurrent switch in the switchgear or, if necessary, by unplugging 
the  plug  of  the  power  cord  from  the  electrical  outlet  in  accordance  with  the  safety  rules 
described  in  these  operating  instructions!  After  disconnecting  the  power  supply,  contact  the 
service department. 

6.4 Sleep mode   

The room sterilizer STERYLIS VS goes into so-called sleep mode when it is connected to the power supply and the main 
switch is  in the OFF position. This is a mode in which the device looks off - the user panel is completely extinguished, no 
errors are signalled and no fans, UV lamps and ozone generators are running. However, the concentration of ozone is 
monitored. Possible exceeding of the safe ozone concentration will be indicated by a warning beacon light and indicated 
on the display in the CONCENTRATION section. 

6.5 Front panel locking mode 

 

The STERYLIS VS room sterilizer is equipped with the possibility of activating a front panel lock. It is intended to block the 
device  in  such  a  way  that  undesirable  persons  cannot  operate  it,  including  in  particular  the  possibility  of  accidentally 
starting a sterilization mode. In lock mode, all buttons on the front panel (except those used to exit the mode) are no 
longer active. 
Lock mode can only be activated when the unit is in standby, silent disinfection, standard disinfection or filtration mode.  
Lock mode features: 

 

After entering and leaving the sleep mode, the lock mode remains active 

 

The mode is active without any time limit 

 

When the device is restarted, the lock mode is always inactive 

 

In the lock mode, the display in the O

STERILIZATION section shows "L0CH" (scrolling with possible error codes 

and messages) 

Summary of Contents for STERYLIS VS-100/450

Page 1: ...MODELS VS 100 450 VS 160 600 VS 300 900 VS 400 1200 VS 600 1500 VS 900 1800 MANUAL USER S PROFESSIONAL ROOM STERILISERS...

Page 2: ...with the ozone concentration limits 23 6 3 4 5 Emergency situations during the sterilization process 24 6 4 Sleep mode 24 6 5 Front panel locking mode 24 6 6 Changing the configuration parameters 25...

Page 3: ...inside the device is high and can cause an electric shock This is especially true for the very high voltage generated by the ozone generators It is dangerous to touch any parts inside the device In th...

Page 4: ...arning icons They are located in the manual on the housing of the device or on other parts of the steriliser and are directly related to it Symbol Meaning Symbol Meaning Important safety instruction B...

Page 5: ...pale cold skin sweat chills accelerated heart rate place them in an anti shock position on their back with their legs raised Despite a specially designed disinfection channel and limiting switch prot...

Page 6: ...pain tearing itching redness avoid direct sunlight cool your eyes with a cloth dampened with cold water or glasses with cooling gel Do not scratch or rub your eyes If you have contact lenses it is ess...

Page 7: ...s disease and increased ailments in people with coexisting respiratory and cardiovascular diseases In very high concentrations it is dangerous to live The effect of ozone on living organisms depending...

Page 8: ...es avoid contact with eyes Wash your hands thoroughly with water and soap after providing medical assistance Inhalation If the airways come into contact with too high a concentration of ozone leave th...

Page 9: ...eir indications are correct Improper measurements of the ozone sensor can lead to life threatening and health threatening effects as described in this chapter The ozone sensor is calibrated by the man...

Page 10: ...dance with EN 136 and EN 14387 standards and in the skin protection equipment in accordance with the personal protective equipment table When the ozoning of the room is complete it must be thoroughly...

Page 11: ...in the room including electronics it is also advisable to move them during disinfection The use of ozone sterilisation ozoning from time to time should not affect the destruction of objects in the roo...

Page 12: ...ria and allergens using a coating applied to the filter medium In addition to particles the active carbon filter retains harmful gases and odours through adsorption In the UV C disinfection mode the U...

Page 13: ...on dose 149 J m2 232 J m2 443 J m2 Ozone generator capacity 30 000 mg h 45 000 mg h 60 000 mg h Net weight 18 kg 20 kg 35 kg in sterilisation mode Name STERYLIS STERYLIS STERYLIS Model VS 400 1200 VS...

Page 14: ...ts with other molecules breaking down their structures The effectiveness of ozone in eliminating odours lies in the fact that as a gas it is able to reach wherever there is air i e all the cracks nook...

Page 15: ...pecially selected type and high power of the UV C light source with a selective emitted wavelength of 253 7 nm and the unique solution of the double quartz filter UV C lamp used in STERYLIS steriliser...

Page 16: ...g manual 6 Safety data sheet ozone 7 UV C safety data sheet 8 Safety data sheet 9 Declarations of conformity in Polish and English 10 Two warning signs WARNING NO ENTRY STERILISATION IN PROGRESS as be...

Page 17: ...front of the device The user panel consists of a main switch controls individual modes of operation failure and warning function keys three segment displays and a graphical display of the ozone diagr...

Page 18: ...ONCENTRATION section segment display indicating the ozone concentration and the and SET buttons Alarm icons for failure of a system of ozone generators ozone sensor or UV C fluorescent lamps Main devi...

Page 19: ...enter the standby mode and then display the corresponding error codes on the display in the O3 STERILIZATION section except for errors F201 and F202 which cause the PRE TEST to stop until the error is...

Page 20: ...PRE TEST function and then it will start to light steadily and the device will start working in the selected mode of operation Confirmation of the disinfection mode is the operation of the front panel...

Page 21: ...where the user can change the parameters of ozoning time and working ozone concentration 6 3 4 1 O3 sterilisation automatic mode To start O3 STERILIZATION in automatic mode press and hold the yellow S...

Page 22: ...igh frequency inactive numbers light up constantly The PROG SET button changes the active digit to the next one on the right After entering the correct code confirm it by pressing the yellow START but...

Page 23: ...During the ozone destruction process the LED backlight panel is activated When the safe concentration 0 1 ppm is reached the ozone safety light will light up After completion of the sterilization pro...

Page 24: ...ower supply of the device should be disconnected as safely as possible preferably by means of an overcurrent switch in the switchgear or if necessary by unplugging the plug of the power cord from the...

Page 25: ...generators 0 1 0 C150 Date setting day of the month 1 31 15 C151 Date setting month 1 12 7 C152 Date setting year 2020 2120 2020 C200 FIRMWARE revision read only FW_REV The individual parameters are c...

Page 26: ...value of the electricity consumed is previewed using the d201 meter 6 8 Other functionality of the device Other remaining functionality of the device is described in subsections 6 8 1 6 8 2 and 6 8 3...

Page 27: ...e sensor error has been detected during previous use of the unit the OZONE SENSOR TEST function can erase this error if the unit will function correctly during this function applies to the following e...

Page 28: ...nd setting the current date The function is used to set the current time on the controller Selecting the SET TIME function displays the currently set time on the display in the OZONING TIME section To...

Page 29: ...1 and 7 2 2 Do not let the filters clog and avoid their excessive operation After the service life of the UV lamps 9 000 operating hours in the steriliser they should be replaced This is related not...

Page 30: ...t air filter cartridge towards the floor 4 Remove the air filter from the cartridge 5 Insert a new air filter into the cartridge making sure that the direction of the arrow on the cartridge and on the...

Page 31: ...of the end point also results in a corresponding error on the display in the O3 STERILIZATION section If the end point violation stops the error is automatically reset The safety limiting switches pr...

Page 32: ...s time has been exceeded will be displayed Note The replacement of UV C fluorescent tubes in STERYLIS sterilisers may only be carried out by the manufacturer s service or another authorised service ce...

Page 33: ...STERILIZATION mode according to the instructions in section 6 3 4 9 Exit the room by closing the door behind you and make sure again that it is tightly sealed and that no one is in the room immediatel...

Page 34: ...vice life of air filters F003 Exceeded service life of ozone generators F004 Exceeded service life of the ozone sensor F005 Exceeded expiry time of the ozone sensor F006 Air filters expiration date ex...

Page 35: ...t blinking PRE TEST function is in progress for standard disinfection mode active errors Yellow pictograms O3 and flashing PRE TEST function for sterilization mode in progress no active errors Yellow...

Page 36: ...use within certain voltage limits Directive 2014 30 EU on the harmonisation of the laws of the Member States relating to electromagnetic compatibility Directive 2011 65 EU on the restriction of the u...

Page 37: ...t with the service department Phone 48801080 257 E mail address serwis emiloo pl When contacting the service department you may be required to provide the data on the unit s nameplate Serial number Da...

Reviews:

Related manuals for STERYLIS VS-100/450