background image

 

 

 

Sterylis  BASIC MANUAL                                                                                                                                                                            

  

| 11

 

 

6.

 

Operation 

The device should be operated in rooms whose parameters are consistent with the technical characteristics of the device. 
In  addition,  they  must  be  operated  in  accordance  with  the  information  contained  in  the  chapters:  "Maintenance"  and 
"Important safety instructions". 

 

6.1 Operation of the device

 

To operate the room steriliser, use the user panel on the front of the device. The construction and operation of the user 
panel is described in the following sections. 

 

 

 
 
 

 
 
 
 

 

 
 
 

 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Handle to transport the device 
 
Cassette filters (with activated carbon) 
 
User panel 

Cassette filters (pre-filters) 
 
Feet (transport wheels for selected versions) 

Summary of Contents for STERYLIS BASIC-100

Page 1: ...MODELS BASIC 100 BASIC 150 BASIC 300 BASIC 400 BASIC 800 BASIC 1200 USER S MANUAL PROFESSIONAL ROOM STERILISERS...

Page 2: ...r disinfection standard mode operating mode 13 6 3 3 UV C air disinfection silent mode operating mode 13 6 3 4 UV C air disinfection automatic mode operating mode 14 6 4 Front panel locking mode 14 6...

Page 3: ...information before you start servicing Before starting service work you should also read the UV C radiation safety data sheet All service work must be carried out in accordance with the instructions f...

Page 4: ...manual please contact the manufacturer directly 7 It is essential that you read the meaning of the following warning icons They are located in the manual on the housing of the device or on other parts...

Page 5: ...n pale cold skin sweat chills accelerated heart rate place them in an anti shock position on their back with their legs raised Despite a specially designed disinfection channel and limiting switch pro...

Page 6: ...of more serious symptoms severe blinking pain foreign body impression in the eye it is necessary to contact an ophthalmologist Until an ophthalmologist has given assistance proceed as with lighter sy...

Page 7: ...n drags the right amount of air through the filters and the UV lamp making the cleaning and disinfection process possible In filter mode the room air is cleaned with a pre filter F1 and an activated c...

Page 8: ...W 45 W Average UV C radiation intensity 127 J m2 150 J m2 257 J m2 Net weight 16 kg 17 kg 31 kg IP degree of protection IP20 IP20 IP20 in UV C air disinfection standard mode Name STERYLIS STERYLIS STE...

Page 9: ...obtain a high quality deodorising effect removing odours from the air Due to the special design of the UV C chamber the device operating in disinfection mode can be used in rooms where people are pre...

Page 10: ...ditional 10 meter long power cord 4 Operating manual 5 Safety data sheet 6 UV C safety data sheet 7 Declarations of conformity in Polish and English if you purchase the device as an option with a long...

Page 11: ...ned in the chapters Maintenance and Important safety instructions 6 1 Operation of the device To operate the room steriliser use the user panel on the front of the device The construction and operatio...

Page 12: ...PERATION TIME section with operation time segmented display Alarm icons for fan or UV C fluorescent lamp system failure UV C section with graphic display User panel switch Operating modes section stan...

Page 13: ...n process press and hold the START button located at the green filtration mode light again 6 3 2 UV C air disinfection standard mode operating mode To start the UV C air disinfection standard mode ope...

Page 14: ...old down the button START in the AUTOMATIC MODE section The automatic mode light will start blinking the device will check the proper functioning of the relevant actuators PRE TEST function and then i...

Page 15: ...ION TIME section To change the value of the selected parameter operate the and buttons in the OPERATION TIME section The selection of the active digit is made by means of the RESET button To confirm t...

Page 16: ...ndby mode Then enter the appropriate code to move to the above mentioned functions Entering the access code The 4 digit code entered is shown on the display under AUTOMATIC MODE The and buttons change...

Page 17: ...will go into standby mode 6 7 2 SET TIME function The function is used to set the current time on the controller When the SET TIME function is selected the number active on the display in the TIMER se...

Page 18: ...s clog and avoid their excessive operation After the service life of the UV lamps 9 000 operating hours in the steriliser they should be replaced This is related not only to the possible burning of th...

Page 19: ...nt air filter cartridge towards the floor 4 Remove the air filter from the cartridge 5 Insert a new air filter into the cartridge making sure that the direction of the arrow on the cartridge and on th...

Page 20: ...in the UV C OPERATION SCHEDULE section If the limit violation ceases the error is automatically deleted The safety limiting switches provide protection in case of an attempt to change the air filters...

Page 21: ...EDULE section If the limit violation ceases the error is automatically deleted The safety limiting switches provide protection in case of an attempt to change the air filters during operation 7 3 Repl...

Page 22: ...on incorrect power consumption E201 Air filter error F1 E202 Air filter error F2 F001 Exceeded service life of the UV C lamp F002 Exceeded service life of air filters F006 Air filters expiration date...

Page 23: ...sinfection mode in progress no active errors Green filtration mode light blinking tESt text blinking PRE TEST function for filtration mode in progress active errors Green silent mode light blinking tE...

Page 24: ...for use within certain voltage limits Directive 2014 30 EU on the harmonisation of the laws of the Member States relating to electromagnetic compatibility Directive 2011 65 EU on the restriction of th...

Page 25: ...tment you may be required to provide the data on the unit s nameplate Serial number Date of manufacture Device name and model The information in this document may be changed by without notice Reproduc...

Reviews:

Related manuals for STERYLIS BASIC-100