PL-550
IQ Radar 160/160 Ex
Page 9
Specifications
Process:
{
material dielectric:
{
ε
r
> 1.8
(For
ε
r
< 3, you should use a wave guide antenna or
still pipe. See Mounting: Waveguide Antenna on page
26 or Mounting: Still pipe or Side pipe on page 27.)
{
temperature:
{
UHMW-PE -40 to 80°C (-40 to 176°F)
{
PTFE -40 to 200°C (-40 to 392°F)
5
{
pressure (vessel):
{
dependant on process connection type and
temperature. Refer to Appendix III on page 69 for
charts.
Approvals (refer to device nameplate):
{
safety:
{
CSA
NRTL/C
, CE, FM
{
radio:
{
BAPT, Industry Canada, FCC
{
hazardous areas:
{
IQ Radar 160:
{
CSA Class I/II, Div.2, Group A, B, C, D, F
(standard)
and G, Class III.
{
CE, CSA
NRTL/C
FM (non-hazardous)
{
IQ Radar 160Ex
6
:
{
Cenelec/Sira
7
, EEx de IIB+H
2
T6 (Note: antenna
may be used in Zone 0 environments).
{
FM (USA) Class I, Div 1, Group A, B, C, D. Class
II/III, Div. 1, Group E, F, G (Class I, Zone 1 IIC
T6)
{
CSA Class I/II, Div. 1, Group B, C, D, E, F, G.
{
CE, CSA
NRTL/C
FM (non-hazardous)
{
Canadian Registration Number (CRN) for pressure fittings
8
:
Ontario, British Columbia, Alberta:
OF6494.512
others pending
{
3A Sanitary
Contact Milltronics for complete and up to date list of approvals.
Programmer (remote keypad):
{
enclosure:
{
general purpose
67 mm w x 100 mm h x 25 mm d
(2.6" w x 4" h x 1" d)
{
ambient temperature:
{
-20 to 50° C (-5 to 122° F)
{
interface:
{
proprietary infrared pulse signal
{
power:
{
9V battery (ANSI/NEDA 1604, PP3 or equivalent)
{
weight:
{
150g (0.3 lb)
6
Cenelec: -20 to 40
°
C (-4 to 104°F), CSA and FM: -20 to 60
°
C (-4 to 140°F) ambient
7
Approved
for
PTFE
material
only
8
All process connections except for the sliding waveguide
Summary of Contents for IQ Radar 160
Page 1: ...33455500 Rev 3 1 IQ RADAR 160 Instruction Manual PL 550 January 2001 Q RADAR I60 ...
Page 10: ...Page 10 IQ Radar 160 160 Ex PL 550 Specifications ...
Page 32: ...Page 32 IQ Radar 160 160 Ex PL 550 Installation ...
Page 35: ...PL 550 IQ Radar 160 160 Ex Page 35 Start Up Display Aiming the Hand Programmer ...
Page 40: ...Page 40 IQ Radar 160 160 Ex PL 550 Start Up ...
Page 52: ...Page 52 IQ Radar 160 160 Ex PL 550 Operation ...
Page 62: ...Page 62 IQ Radar 160 160 Ex PL 550 Parameter Descriptions ...
Page 66: ...Page 66 IQ Radar 160 160 Ex PL 550 Appendix I ...
Page 68: ...Page 68 IQ Radar 160 160 Ex PL 550 Appendix II ...
Page 74: ...Page 74 IQ Radar 160 160 Ex PL 550 Appendix III ...