Page 28
IQ Radar 160/160 Ex
PL-550
Installation
Mounting: Horn with Waveguide Extensions
In applications where the stand pipe is too
long and the diameter is too small (such as a
stand pipe that is 100 mm (4”) in diameter
and 460 mm (18”) in length), the rod antenna
is not suitable due to stand pipe interference.
In this case, use the waveguide / horn
combination. Waveguide extensions are
available in custom lengths.
If the horn diameter is too large for the stand
pipe opening, you need to insert it from inside
the vessel. The horn must be connected to
the IQ160 process flange.
Note:
The IQ160 maximum range of 15m is reduced by [0.64 x waveguide
length]. Blanking and offset parameters will be set by Milltronics, see the
device tag for values.
Mounting: Sanitary Mounting
There are two common sanitary mounting options; the 2”, 3” and 4” tri-clamp
with rod antenna and the sanitary 4” horn antenna.
Wetted Parts:
PTFE or UHMW-PE only.
Summary of Contents for IQ Radar 160
Page 1: ...33455500 Rev 3 1 IQ RADAR 160 Instruction Manual PL 550 January 2001 Q RADAR I60 ...
Page 10: ...Page 10 IQ Radar 160 160 Ex PL 550 Specifications ...
Page 32: ...Page 32 IQ Radar 160 160 Ex PL 550 Installation ...
Page 35: ...PL 550 IQ Radar 160 160 Ex Page 35 Start Up Display Aiming the Hand Programmer ...
Page 40: ...Page 40 IQ Radar 160 160 Ex PL 550 Start Up ...
Page 52: ...Page 52 IQ Radar 160 160 Ex PL 550 Operation ...
Page 62: ...Page 62 IQ Radar 160 160 Ex PL 550 Parameter Descriptions ...
Page 66: ...Page 66 IQ Radar 160 160 Ex PL 550 Appendix I ...
Page 68: ...Page 68 IQ Radar 160 160 Ex PL 550 Appendix II ...
Page 74: ...Page 74 IQ Radar 160 160 Ex PL 550 Appendix III ...