7ML19981AN02
AiRanger XPL Plus – INSTRUCTION MANUAL
Page 7
mm
mmm
Inst
allat
ion
Installation
Installation shall only be performed by qualified personnel, and in accordance with
local governing regulations.
The following procedure applies to all XPL Plus level monitor installations. See
Technical
Reference
Application Examples for additional installation requirements. Also, refer to
the instruction manuals of all other equipment connected to the XPL Plus for additional
installation instructions.
AiRanger XPL Plus
Installation should only be performed by qualified personnel, and in accordance with all
governing regulations.
Location
Inspect all potential mounting locations. Choose a location suited to the XPL Plus
polycarbonate enclosure and the following location recommendations.
The ideal XPL Plus mounting location is where the
1. Ambient temperature is always within -20 to 50 °C (-5 to 122 °F).
2. XPL Plus front cover has sufficient room to swing open.
3. Operator and electronics are not exposed to the weather.
4. Cable length requirements are minimal.
5. Mounting surface is free from vibration.
Avoid mounting locations where the XPL Plus is
1. Exposed to direct sunlight. (Otherwise, provide a sun shield.)
2. Close to high voltage or current runs, contactors, or SCR control drives.
Cable / Conduit Entry Requirements
Determine the number of enclosure cable / conduit entries required for
• Transducers
• TS-3 temperature sensors (if used)
• BIC-II (if used)
• SAM-20 (if used)
• AO-10 (if used)
• Synchronization (see Interconnection-Level System Synchronization)
• Power
Note:
Transducer cables must be run in a grounded metal conduit, separate
from other wiring, (except TS-3 temperature sensor wiring, if applicable).
Summary of Contents for airanger XPL plus
Page 1: ...Instruction Manual March 2005 XPL PLUS airanger ...
Page 10: ...Page 6 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Page 26: ...Page 22 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Page 82: ...Page 78 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Page 106: ...Page 102 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Page 113: ......