Millipore Centriprep 50k User Manual Download Page 1

User Guide

Centriprep

®

Centrifugal Filter Devices

 

Introduction

Centriprep® centrifugal filter devices are disposable ultrafiltration devices used for purifying, 

concentrating, desalting, and filtering biological samples in the 2–15 mL volume range. These ready-

to-use ultrafiltration devices are designed for operation in most centrifuges that can accommodate 

50 mL centrifuge tubes. They are easy to use and offer a high flow rate. The Centriprep® device consists 

of a sample container with a twist-lock cap, a filtrate collector containing a low adsorptive Ultracel® 

regenerated cellulose membrane, and an air-seal cap for sample isolation.
The Centriprep® device design allows filtration and fine particle sedimentation to occur at the same time 

without clogging the membrane. With a Centriprep® 50K device, a 15 mL sample can be concentrated to 

0.6 mL in just 20 minutes and concentration of smaller volumes takes even less time. High concentration 

factors are achieved because of the membrane’s low adsorptivity, and greater than 90% recovery of retained 

macromolecular solutes is typical. A built-in deadstop provides a final concentrated sample volume of 

0.5–0.6 mL.
The Centriprep® product line includes 4 different molecular weight cutoffs (MWCO):

 

● Centriprep® 3K device — 3,000 MWCO (yellow membrane support)

 

● Centriprep® 10K device — 10,000 MWCO (green membrane support)

 

● Centriprep® 30K device — 30,000 MWCO (colorless membrane support)

 

● Centriprep® 50K device — 50,000 MWCO (pink membrane support)

NOTE: 

Centriprep® 10K and 30K devices are the only Centriprep® devices intended for in vitro diagnostic 

use. For information on these devices, go to www.millipore.com/centrifugal_ivd_userguide.

Applications

Centriprep® 3K centrifugal filter devices:

 

● Concentrating and desalting oligonucleotides, peptides, growth factors, and small proteins.

Centriprep® 10K, 30K, and 50K centrifugal filter devices:

 

● Concentrating and desalting column eluates and gradient fractions containing proteins, enzymes,  

and antibodies

 

● Recovering biomolecules from cell culture supernatants, lysates, extracts, or other biological samples

 

● Purifying low-molecular-weight components (e.g., amino acids and antibiotics) from physiological 

fluids, cell culture media, or fermentation broths.

Centriprep® Device Components

Air Seal Cap

Twist-Lock Cap

Vent Groove

Filtrate Collector Shoulder

Filtrate Collector

Membrane Support

Sample Container

Fill Line

Required Equipment

Centriprep® devices can be spun in a variable speed centrifuge with either a swinging-bucket or fixed-

angle rotor and carrier that can properly accommodate 50 mL centrifuge tubes. The centrifuge must be 

capable of 500–3,000 × g. For a listing of compatible centrifuges and rotors, see www.millipore.com/

rotorguide.

Limitations

With fixed-angle rotors, polarization control may be adversely affected at low g-forces (below 2,000 × g). 

Depending on sample composition and solute concentration, filtrate flow may be reduced. When using 

either type of centrifuge, do not exceed the following limits because excessive g-force may result in 

leakage or damage to the device.

Device

Maximum Centrifugal Force

Centriprep® 3K

3,000 × g

Centriprep® 50K

1,500 × g

Device Storage and Shelf Life

For best results, store Centriprep® devices at 15–30 °C. Performance may be adversely affected if the 

devices are stored at extreme temperatures (i.e., below 0 °C or above 40 °C). Shelf life is three years from 

date of manufacture.

Rinsing Before Use

The membranes in Centriprep® centrifugal filter devices contain trace amounts of glycerine. If this inter-

feres with analysis, fill the device with approximately 15 mL of buffer or deionized water and centrifuge 

until the first equilibration point is reached. Decant filtrate and concentrate. Repeat as necessary until no 

more interference is noted. Once wet, membranes must be kept wet to remain functional.

Membrane and Device Appearance

 

● Occasionally, an unused Centriprep® device may have a few drops of moisture in the filtrate collector 

or sample container. This is due to condensation of moisture in the membrane and does not affect 

performance.

 

● After centrifugation, a small amount of filtrate (approximately 0.5 mL) may remain in the area 

between the membrane and the membrane support of the Centriprep® device. This is normal and does 

not affect device operation or performance.

 

● The Ultracel® membrane used in the Centriprep® centrifugal filter devices may appear slightly wrinkled 

after spinning.

Centrifugal Filter Device Reuse

Centriprep® devices are intended for single use only. Attempts to clean the device may damage the 

membrane surface or the seal, possibly compromising the unit’s integrity and thereby affecting 

performance.

Principle of Operation

To begin operation, the sample container’s twist-lock cap is loosened and the filtrate collector is removed. 

Sample is added, then the filtrate collector is carefully reinserted back into the sample container (displacing 

solution) and the cap locked, sealing the device. The assembled centrifugal filter device is centrifuged at 

1,000–3,000 × g, depending on the membrane MWCO used.
Immersing the filtrate collector in the sample solution creates a slight hydrostatic pressure differential 

that exerts an upward buoyancy force on the membrane at the filtrate collector bottom. By itself, the 

hydrostatic pressure exerted by the displaced solution is too weak to produce ultrafiltration. Centrifugation 

increases this pressure, forcing low molecular weight materials and solvent through the membrane into 

the filtrate collector. Solutes with molecular weights above the membrane MWCO remain in the sample 

container and become increasingly concentrated as the operation continues. Ultrafiltration occurs in the 

direction opposite the centrifugal force vector.

Filtr

ation

Centrifugal Force

Swinging-Bucket Rotor

Fixed-Angle Rotor

Filtration

Polarization 

Layer

The pressure differential created by the raised solution level is important in centrifugal filter device 

operation. During centrifugation, the sample solution meniscus falls as the filtrate meniscus rises. In the 

process, the filtrate collector loses its buoyancy and sinks to the bottom of the sample container, permitting 

maximum filtrate collection. Eventually, an equilibrium is reached where the menisci are at equal heights. 

Filtration then stops, since the hydrostatic pressure difference is now zero.
If further concentration is required, decant the filtrate before spinning the device for a second time. 

Decanting re-establishes the pressure differential between the sample meniscus and the filtrate meniscus, 

allowing filtration to resume. Filtration continues until a new equilibrium point is achieved. Note that the 

pressure differential is constantly changing during operation as the menisci rise and fall.
Centriprep® centrifugal filter devices are designed to maintain high flow rates by minimizing solute 

build-up on the membrane during operation. Centrifugal force causes dense materials to sink away from 

the membrane to the sample container bottom. Because the Centriprep® membrane “floats” above this 

polarization layer, the full surface of the membrane is used at all times, and filtrate flow is unrestricted. 

This is particularly beneficial when working with suspensions, which can easily clog or foul a membrane.

How to Use Centriprep® Centrifugal Filter Devices

Spin to  

concentrate

Spin to  

concentrate

Disassemble device and  

add sample to fill line.

Reassemble and  

place in centrifuge.

Spin until solution 

levels have equalized.

Orient vent groove up and  

decant filtrate. Spin again.

Disassemble and  

recover concentrate.

1.  Turn the twist-lock cap on the sample container counterclockwise, then slide the filtrate collector/cap 

assembly out and set it aside.

CAUTION: 

Take care not to touch, scratch, or damage the membrane on the bottom of the filtrate 

collector during disassembly, filling, and re-assembly of the device.

2.  Add solution to the sample container. The line on the side of the container marks the maximum fill 

volume of 15 mL. Do not exceed this volume.

NOTE: 

For solutions with particulate material, such as cell suspensions, a starting volume over 5 mL 

will result in reduced flow rates. For best results when working with a sample having over 10% 

solids, limit the initial sample volume to 5 mL or less.

3.  Seat the twist-lock cap fully onto the shoulder of filtrate collector. If necessary, slide the cap 

downward until it stops at the shoulder.

4.  Carefully insert the capped filtrate collector into the sample container, gently pushing down so the 

collector displaces the solution. Turn the twist-lock cap clockwise to seal the sample container. Finally, 

make sure the air-seal cap is snug on the twist-lock cap.

5.  Insert the assembled Centriprep® device into the centrifuge and counterbalance with a similar device.

CAUTION: 

Inspect swinging-bucket rotors for proper clearance before centrifuging. Any obstruction 

may result in damage to the Centriprep® device and possible loss of sample. For added 

clearance, the Centriprep® device may be used in adaptors without rubber cushions.

6.  Spin the Centriprep® device at the appropriate g-force until the fluid levels inside and outside the 

filtrate collector equilibrate. For concentration applications, see the guidelines for achieving various 

concentration volumes outlined in the following “Concentration Times for Dilute Protein Solutions” 

section.

7.  If further concentration is required after equilibration, remove the device from the centrifuge and take 

off the air-seal cap. With the vent groove oriented upward, decant the filtrate; replace the cap and 

spin the device a second time. For filtration applications, reserve the filtrate.

8.  After the second spin, decant the remaining filtrate. If further concentration is desired, spin the device 

again; otherwise, proceed to step 9.

9.  Loosen the twist-lock cap (turn counterclockwise) and remove the filtrate collector. Withdraw the 

sample using a pipette, or pour the concentrate into a suitable container.

Typical Concentration Times for Dilute Protein Solutions

Starting Volume  

in Device

Spin Time (min) and Concentrate Volume (mL)

Centriprep® 3K

First Spin

Second Spin

Third Spin

min

mL

min

mL

min

mL

5 mL

50

1.2

10

0.8

10

0.5

10 mL

65

2.0

10

1.0

10

0.5

15 mL

95

3.1

35

1.2

10

0.6

Centriprep® 10K

First Spin

Second Spin

Third Spin

min

mL

min

mL

min

mL

5 mL

15

1.3

10

0.6

N/A

10 mL

30

2.1

10

1.0

5

0.5

15 mL

40

3.2

10

1.2

5

0.5

Centriprep® 30K

First Spin

Second Spin

Third Spin

min

mL

min

mL

min

mL

5 mL

5

1.3

5

0.8

5

0.5

10 mL

10

2.4

5

1.2

5

0.4

15 mL

15

3.2

10

1.1

5

0.5

Centriprep® 50K

First Spin

Second Spin

Third Spin

min

mL

min

mL

min

mL

5 mL

5

1.4

5

1.0

5

0.6

10 mL

10

2.2

5

1.1

5

0.6

15 mL

10

3.4

5

1.3

5

0.6

Spin Conditions: Swinging-bucket rotor, 3,000 × g for 3K and 10K, and 1,500 for 30K and 50K, 25 °C  

Protein marker used: Cytochrome c (0.25 mg/mL) for 3K, and bovine serum albumin (1 mg/mL) for 10K, 30K, and 50K. 

N/A = not applicable, third spin unnecessary.

Performance

Flow rates and recovery characteristics of Centriprep® centrifugal filter devices have been tested with 

several well-known solutes. These results, which are summarized in the following sections, may be used to 

estimate performance with other solutes.

Relative centrifugal force (RCF) is measured at the filtrate collector base and is calculated as follows:
RCF = 1.118 × 10

-5

 × radius × (RPM)

2

Radius = distance in centimeters, measured from the  

 center of rotation to the base of the filtrate collector

RPM = revolutions per minute.

NOTE: 

RCF and RPM are not the same (i.e., 3,000 × g is NOT equivalent to 3,000 RPM). Check the 

centrifuge operating guide for instructions on converting g-force to RPM, or use the formula 

above.

Flow Rate

Filtration rate is affected by several operating parameters, including sample concentration, starting 

volume, relative centrifugal force, type of rotor used, membrane MWCO, and temperature. When 

concentrating dilute solutions, swinging-bucket and fixed-angle rotors yield comparable flow rates. 
For solutions containing over 10% solids, a starting volume greater than 5 mL typically results in reduced 

flow rates. A volume of 5 mL or less produces the fastest flow rates because this limits the amount of solids 

that can pack onto and blind the membrane.
Spin times must be lengthened when working at low temperatures. At 4 °C, flow rates are approximately 

1.5 times slower than at 25 °C. Viscous solutions may also require longer spin times. For example, a  

1 mg/mL BSA solution in 50% glycerine would take five times longer to concentrate than the identical 

protein in an aqueous buffer solution, using a Centriprep® 10K device.

Retention and Recovery

The anisotropic, hydrophilic membranes in Centriprep® centrifugal filter devices are characterized by a 

molecular weight cutoff; i.e., their ability to reject molecules above a specified molecular weight.
Because it is an easy parameter to work with, molecular weight is used in rating membrane permeability 

and selectivity based on results achieved with well-characterized proteins.
However, molecular weight may not be valid for all solutes. For instance, retentivity may be greater for 

a globular molecule than for a linear molecule having the same nominal molecular weight. The effects 

of pH and ionic strength on a molecule’s dimensions and behavior in a particular solution must also be 

considered when assessing solute retentivity.
Low solute recovery in the concentrate may indicate possible adsorptive losses and/or solute passage 

through the membrane. Adsorptive losses depend on solute concentration, the nature of solute 

(hydrophobic or hydrophilic), temperature, time of contact with component surfaces, sample composition, 

and pH.

Typical Concentrate Recovery

Solute & Starting Concentration

Nominal 

Molecular 

Weight

 

Centriprep® % Concentrate Recovery

3K

10K

30K

50K

IgG Fr. II (1 mg/mL)

156,000

99

Bovine serum albumin (1 mg/mL)

67,000

97

98

93

Ovalbumin (1 mg/mL)

45,000

95

90

α

-Chymotrypsinogen (1 mg/mL)

25,000

94

93

Cytochrome c (0.25 mg/mL)

12,400

96

96

8

4

Vitamin B12 (0.2 mg/mL)

1,355

12

5

15 mL starting volume, 25 °C, swinging-bucket rotor
Results based on recommended spin times outlined in the “Concentration Times for Dilute Protein 

Solutions” section

Typical Protein Recovery, Centriprep® 30K Device

Concentration

% Concentrate Recovery ± 1 SD

100 µg/mL

98.2 ± 1.7

1,000 µg/mL

96.6 ± 0.2

15 mL starting volume, 25 °C

,

 swinging-bucket rotor, 1,500 × g 

Bovine serum albumin

two 25-minute spins

Desalting

For desalting or solvent exchange, the sample is concentrated until the equilibration point is reached. 

Filtrate is discarded, then the sample is reconstituted to the original volume by adding an appropriate 

solvent and vortex mixing. The sample is then concentrated and reconstituted once more. This process is 

repeated until the concentration of the contaminating microsolute is sufficiently reduced.
The following theoretical example illustrates the process of sample desalting:

Process Step

Sample Volume Protein Concentration NaCl Concentration

Initial sample

15 mL

1 mg/mL

500 mM

Spin to concentrate

3 mL

5 mg/mL

500 mM

Reconstitute back to 15 mL  

with 10 mM NaCl

15 mL

1 mg/mL 

108 mM

Spin to concentrate/ 

desalt #1

3 mL

5 mg/mL

108 mM

Reconstitute back to 15 mL  

with 10 mM NaCl

15 mL

1 mg/mL

29.6 mM

Spin to concentrate/ 

desalt #2

3 mL

5 mg/mL

29.6 mM

Reconstitute back to 15 mL  

with 10 mM NaCl

15 mL

1 mg/mL

13.9 mM

Spin to concentrate/ 

desalt #3

3 mL

5 mg/mL

13.9 mM

Centriprep® 3K and 50K devices for research  
use only; not for use in diagnostic procedures

Performance, 

continued

PR04321_RevA.indd   1

8/29/2013   1:34:59 PM

Reviews: