Millipore Sigma Pellicon 3 Cassettes Installation And User Manual Download Page 11

Pellicon

®

 3 Cassettes Installation and User Guide 

11

5.  Slowly raise the air pressure to the 

recommended test value and wait 

5 minutes to purge residual water 

in the permeate line. Permeate line 

must be completey emptied before 

measuring air flow. Do not exceed 

the recommended air pressure 

as this may displace water from 

the membrane pores and result in 

excessively high air flow (a false 

failure). Rewet the membrane if this 

occurs. 

6.  After the diffusion air flow pressure 

stabilizes, measure and record the 

air pressure, temperature and the 

air flow rate exiting the permeate 

line. The air flow rate may be 

measured with an air flow meter or 

by measuring the air displaced into a 

submerged and inverted volumetric 

cylinder as shown in the figure above. 

7.  Compare the measured air flow 

rate to the specified flow value on 

the Certificate of Quality included 

with the cassette. If the measured 

flow rate exceeds the specified flow 

value, refer to the Troubleshooting 

section of this guide.

Measurement of 

Normalized Water 

Permeability

The normalized water permeability (NWP) 

for Pellicon

® 

 3 Cassettes should be 

established prior to the first product contact 

of each filter. New membranes should be 

cleaned, flushed and integrity tested before 

measuring NWP. The NWP measured at 

this point is used as a benchmark against 

subsequent NWP measurements to 

determine cleaning efficacy.
1.  Close the tank drain valve.
2.  Fill the tank with filtered, deionized 

water, water for injection, or 

reverse osmosis permeate. The 

flush water must be extremely pure 

to avoid fouling the membranes 

or introducing other contaminants 

into the system. The NWP may 

be measured with solutions other 

than water (i.e. storage solution 

or buffer) as long as the same 

conditions and solution are used for 

each measurement.

3.  Fully open the retentate (and 

permeate valve if present) and 

direct both the retentate and 

permeate lines back to the tank. 

4.  Open the tank drain valve
5.  Turn on the feed pump and pump 

water into the feed port of the 

Pellicon

® 

 holder. The recommended 

feed flow rate for Pellicon

® 

 A or C 

screens is 4 to 6 L/min/m².  (The 

same conditions MUST be used each 

time NWP is measured in order to 

ensure accuracy.) 

6.  Recirculate the water for 10 minutes. 

Ensure that the pressure and the 

temperature conditions are stable. 

7.  Record the feed and permeate flow 

rate, feed, retentate, and permeate 

pressures, and the temperature of 

the water. 

8.  At the end of the recirculation, fully 

open the retentate and filtrate lines, 

turn off the feed pump, and drain 

the retentate and filtrate piping. 

9.  Calculate the NWP:  

 

NWP= 

R•F/[A(P

in

 + P

out

/2)-P

p

]

 

 

These units yield  

LMH/bar [liters/m

2

 •hours•bar] 

 

Calculate: 

R = Permeate Flow Rate in L/hour  

P in = Feed Pressure in bar 

P out = Retentate Pressure in bar  

T = Water Temperatures in °C 

Pp = Permeate Pressure (if non-

zero) in bar  

A = Total Filter area in m

2

 

F = Temperature correction factor 

from table below.

The acceptance criterion for cleaning 

efficacy as measured by NWP is 

membrane and application specific, 

and may vary between plants. Key 

Summary of Contents for Pellicon 3 Cassettes

Page 1: ...ale Holder Installation 4 Hydraulic Process Scale Holder Installation 5 Maintenance 7 Flushing and Cleaning 7 Flushing 8 Cleaning 8 Sanitization 9 Depyrogenation 10 Integrity Testing 10 Measurement of...

Page 2: ...TE Pellicon 3 88 cm2 and 0 11 m2 Cassettes must be installed in a Pellicon Cassette Holder Catalog Number XX42PMINI Holder Installation 1 Loosen the nuts on the tie rods of the holder and remove the n...

Page 3: ...wipe 4 Slide the manifold adapter plate onto the holder against the holder s manifold plate with the solid light grey side labeled Manifold Side towards the manifold and the wings resting on the tie...

Page 4: ...ary Tri Clover connection Pellicon Cassette hex nut washer stainless steel bushing handle PERMEATE RETENTATE FEED Pellicon10_2_06_2 eps Gasket Plate manifold adapter plate nut PERMEATE R E T E N T A T...

Page 5: ...wrench and 15 16 in deep hex socket turn each nut 1 8 turn in the sequence shown in Figure 4 When the wrench clicks this indicates that the set force has been reached DO NOT OVER TORQUE Evenness of to...

Page 6: ...d tight 4 Turn on the air supply to the Hydraulic Power Unit A minimum of 80 psig air pressure should be available Check the air pressure on the pressure gauge PI 1 on Hydraulic Power Unit front panel...

Page 7: ...ut replacement of nuts and washers should be considered after every ten installations to maintain high sealing force in the holder 7 When new nuts do not spin freely refurbish the threaded rods with a...

Page 8: ...se the tank drain valve V1 and the tank valve V2 2 Fill the tank with the flushing solution 3 Fully open the retentate valve V4 and permeate valve V5 if present and direct both the retentate and filtr...

Page 9: ...iofilms Biopolymer Proteins and Polyphenolic Membrane Cleaning Agent Concentration Temperature C pH Time min Ultracel NaOH 0 1 N 0 5 N 25 50 13 13 7 30 60 Biomax NaOH 0 1 N 0 5 N 25 50 13 13 7 30 60 N...

Page 10: ...c Acid 0 1M 30 50 1 30 NaOH 0 1N 25 50 13 30 Ultracel NaOH 0 1N 25 50 13 30 Use milder conditions if effectiveness of agent is still observed Better membrane life is expected with lower concentration...

Page 11: ...l the tank with filtered deionized water water for injection or reverse osmosis permeate The flush water must be extremely pure to avoid fouling the membranes or introducing other contaminants into th...

Page 12: ...0 40 0 734 75 2 24 1 023 46 4 8 1 551 102 2 39 0 748 73 4 23 1 047 44 6 7 1 598 100 4 38 0 762 71 6 22 1 072 42 8 6 1 648 98 6 37 0 777 69 8 21 1 098 41 0 5 1 699 Based on Water Fluidity Relative to...

Page 13: ...ess compression and ensures that the units will not leak externally NOTE The storage agent used must comply with applicable local regulations Recommended Storage Solutions for Pellicon 3 Cassette Filt...

Page 14: ...completely flushed Flush system and cassette System and cassette not completely cleaned Clean system and cassette Symptom Possible Cause Remedy High NWP Damage to membrane or cassette Integrity test...

Page 15: ...xed below fluid air interface Vortex in feed container Add baffle to break vortex or reduce mixing speed Retentate splashing Fix retentate line below fluid in feed air interface container Pump cavitat...

Page 16: ...atory matters to the best of our knowledge and ability but without obligation or liability Existing laws and regulations are to be observed in all cases by our customers This also applies in respect t...

Reviews: