92
Cleaning the Inlet Strainer
Purpose
•
The purpose of the Inlet Strainer is to prevent a large particle from entering the
System.
•
If the Inlet Strainer becomes clogged, then feed water does not flow freely to the
System.
•
Cleaning the Inlet Strainer removes any trapped debris.
When
The Inlet Strainer should be cleaned when the following Alert message is displayed.
•
Alert message = EXAMINE INLET STRAINER
The Inlet Strainer should also be cleaned whenever you suspect it is clogged.
Procedure
Follow the steps below to clean the Inlet Strainer.
Step
Action
1
Go to STANDBY Mode.
2
Shut off the feed water supply.
3
Unscrew the Inlet Strainer from the feed water supply.
4
Detach the tubing on the other end of the Inlet Strainer.
5
Flush water backwards through the Inlet Strainer.
6
Apply 3 to 4 turns of new white tape to the threads of the feed
water pipe.
7
Screw the Inlet Strainer back onto the feed water pipe.
8
Attach the tubing to the other end of the Inlet Strainer.
9
Open the feed water supply valve.
10
Go to READY Mode.
Continued on next page