Summary of Alert messages
Alert messages
LCD message
What it means
CALIBRATION VISIT OVERDUE XX
DAYS
•
The System has determined that a
Calibration Visit is overdue.
•
Contact Millipore SAS.
CHECK ASM UV LAMP
•
The ASM UV Lamp is not turning on.
•
Contact Millipore SAS.
CHECK UV 185 NM LAMP
•
The UV 185 nm Lamp is not turning on.
•
Contact Millipore SAS.
CHECK UV 254 NM LAMP
•
The UV 254 nm Lamp is not turning on.
•
Contact Millipore SAS.
EXAMINE INLET STRAINER
•
The System has determined that it is
time to clean the Inlet Strainer.
•
Clean the Inlet Strainer and reset the
message.
LOW RO PUMP PRESSURE
•
The Water System has determined that
the RO Pump Pressure is below
specification.
•
Contact Millipore SAS.
NEXT CALIBRATION VISIT IN XX
DAYS
•
The System is prompting you that a
Calibration Visit should be scheduled.
•
Contact Millipore SAS.
NEXT QUALIFICATION VISIT IN XX
DAYS
•
The System is prompting you that a
Qualification Visit should be scheduled.
•
Contact Millipore SAS.
NEXT SERVICE VISIT IN XX DAYS
•
The System is prompting you that a
Service Visit should be scheduled.
•
Contact Millipore SAS.
NO RESPONSE FROM DHCP
SERVER
•
Contact your network administrator.
•
Restart the System.
PERFORM RO CL2 CLEANING
•
The Water System has determined that
it is time to perform a RO Cl2 cleaning.
•
Start a RO Cl2 cleaning. The timer is
reset automatically.