101
Summary of Alarm messages
Alarm stop
messages
LCD message
What it means
FLOW AUTO STOP
•
This alarm stops the Milli-Q® portion
of the system.
•
A safety feature of the Milli-Q® Direct
system has automatically stopped
dispensing water after 40 minutes to
avoid risks of overflow.
•
Push the POD Unit Plunger all the
way down and release.
•
This resets the dispenser timer and
makes the POD Unit available for
dispensing.
INCORRECT PROGARD CARTRIDGE
•
This alarm stops the RO portion of
the system.
•
The System does not recognize the
type of Progard® Cartridge being
installed.
•
Contact Millipore SAS.
INCORRECT Q-PAK PACK
•
This alarm stops the Milli-Q® portion
of the system.
•
The System does not recognize the
type of Q-Pak® Pack being installed.
•
Contact Millipore SAS.
LOW FEED WATER PRESSURE
•
This alarm stops the RO portion of
the system.
•
Check Feed water pressure and
rectify.
•
Go to STANDBY Mode and go to
READY Mode to release any trapped
air in the Water System.
•
Contact Millipore SAS if the problem
persists.
PERMEATE C > SP
•
This alarm stops the RO portion of
the system.
•
The Permeate conductivity is above
the set point.
•
Contact Millipore SAS.
POD LOCKED
•
This alarm stops the Milli-Q® portion
of the system.
•
The POD Unit was left in the open
position.
•
Push the Plunger all the way down
and release.
Continued on next page