HistoDream EW
MM164-001-Operator Manual
31
2.12.2. Temperature setting
On the operating screen, press the TEMP SET key
, and the following screen will open:
SET TEMPERATURE WAX
60°C
Note that the temperature units of measure (Celsius/Fahrenheit) will
depend on the settings as described under chapter 2.10 Operating settings.
Use the UP or DOWN keys to change the value, if required, and press
ENTER to confirm your selection.
The temperature values for other HistoDream EW zones can be then changed in the order shown below.
Repeat the same procedure for each of the other zones.
The recommended values for each zone (factory default values) are:
ZONE
°Celsius
°Fahrenheit
WAX
60
140
AREA
60
140
VALVE (indirectly heated by the tube)
62
144
FRONT CHAMBER (cassettes)
62
144
REAR CHAMBER (moulds)
62
144
DUCT
This is a tube conveying wax from the
storage cavity to the solenoid valve
62
144
COLD MODULE
-10
14
AUXILIARY TWEEZERS
62
144
2.12.3. Temperatures with a standalone cold module
If the cold module operates as a standalone unit, the temperature is adjusted via the potentiometer located
on the back of the instrument.
The following list shows the temperatures of the cold module based on the potentiometer knob position:
Knob
position
Temperature
in ºC (± 1ºC)
Temperature
in ºF (± 2ºF)
1
0
32
2
0
32
3
-1.2
30
4
-2.7
27
5
-4.0
25
6
-5.4
22
7
-6.8
20
8
-8.2
17
9
-9.6
15
10
-11.4
11.5
Please note that, for safety reasons, some values are internally limited by the unit. This is
why no temperature values exceeding certain limits can be selected.