When must
the device be
serviced?
What must be serviced?
Who must
service the
device?
Before every
workout
Check the device for signs of damage:
n
Control unit (Fig. 62/1)
n
Electrical lines and plug connectors (Fig. 62/2)
n
i-body
®
Replace damaged parts immediately with genuine spare parts of the
manufacturer
.
Do not use the device if there are signs of damage.
Trainer
Every
24 months
Check the electrical system.
Licensed electri-
cian
13.2 Maintaining the electrodes
To ensure optimum function, it is recommended to replace the
electrodes after 12 months of use.
13.3 Handling error messages
If a malfunction was identified that affects the operational readi-
ness, the active training session (voltage generation) is blocked
and an error code is displayed in the status line as well as in the
LED line. Configurative operation functions remain enabled.
Notify the manufacturer and the relevant authority of the member
state in which the operator is located of all major incidents that
occur in connection with the device.
Proceed as follows to fix the error:
1.
Remember or write down the error code.
2.
Switch off the device with the on/off switch.
3.
Switch on the device with the on/off switch.
It takes approx. 17 seconds after the device has been
switched on before it is read for use.
4.
Check whether the error is displayed again.
n
If the error is no longer displayed, continue with the
workout.
n
If the error is displayed again or another error code is dis-
played, notify the manufacturer or Customer Service and
report the error codes (
).
Ensuring optimum function
2021-07-28
miha-bodytec m.ove EMS training device
112