en - Maintenance
56
Remove the drip tray.
The drip tray can be reprocessed by machine in a suitable load car-
rier.
Checking the locking system
The transfer trolley has a locking system to prevent mobile units from
accidentally tipping over. The system is not released until a transfer
trolley is correctly docked onto the conveyor system or cleaning ma-
chine and locked in place.
The locking mechanism is at the front of the transfer trolley. If the
transfer trolley is not docked, the conveyor system will not be able to
push the mobile unit.
The locking system must be checked regularly:
Push the transfer trolley against the cleaning machine. Check that
the trolley and chamber rail have a common axis.
Notify the Miele Customer Service Department if the locking
function is not present. If this is the case, do not use the transfer
trolley.