Maintenance measures
122
Routine checks
Before the start of each working day, the operator must conduct a
series of routine checks. A routine checklist is supplied with the ma-
chine.
The following items must be checked:
- all filters in the wash cabinet
- spray arms in the machine and in any mobile units, modules, or
baskets
- the wash cabinet and the door seal
- dispensing systems
- mobile units, baskets, modules, and inserts
- the filters in the load carriers
Cleaning the filters in the wash cabinet
The filters in the floor of the wash cabinet prevent coarse soiling from
coming into contact with the circulation system. Filters can become
blocked by soiling, so they need to be checked every day and
cleaned as necessary.
Risk of damage due to blocked waterways.
If the filters are not inserted, dirt particles will end up in the ma-
chine water circuit. The dirt particles may block the nozzles and
valves.
Only start a program if the filters are inserted.
Check that the filters are positioned correctly when you reinsert
them after cleaning.
In the controls, it is possible to set a cleaning interval for the filters in
the wash chamber, see “Settings
/Filter maintenance”.
The cleaning interval is not a substitute for the daily routine check of
the filters in the wash chamber!
Danger of injury from glass shards, needles etc. which are re-
tained in the filter.