        Handpiece Maintenance Instructions

After each use:

1.

 Disconnect handpiece from coupler.

2.

 Remove bur from chuck.

3.

 Scrub and rinse under warm tap water with a soft 

  brush or sponge making sure to keep head up to 

  prevent excess water from getting into the back of the handpiece. Make 

sure fiber optic port near head of handpiece is scrubbed well. 

4.

 Dry handpiece with a gauze pad or paper towel.

5.

 Insert handpiece into sterilization bag and load into 

autoclave.   Midwest recommends using sterilization 

bags with at least one paper side.  Handpieces should 

be processed in the autoclave for 15 minutes at 

sterilization temperature of 135˚C (275˚F). For cassette 

chamber saturated steam devices, handpieces should be 

processed unbagged for 3.5 minutes at sterilization tem-

perature of 135˚C (275˚F). STEAM STERILIZE ONLY. Do 

not use other means such as dry heat or chemical.

6.

 The Stylus EasyCare handpiece is now ready t

o use.

Weekly or as-needed maintenance:

If the push button gets stiff:

• Place 1-drop of Midwest Plus Handpiece Lubricant 

(#380130) into chuck.

To clean water spray holes:

• Run the cleaning wire (supplied with handpiece) back 

and forth in the four outer water spray holes and in the 

four inner air spray holes to loosen and remove any 

mineral build-up.

To remove debris from fiber optics:

 • Gently rub both ends of the fiber optic element with an 

alchol soaked cotton swab.

CAUTION:

 

 Other than weekly chuck lubrication as described, 

DO NOT

 

apply or

spray cleaners or lubricants into your Midwest Stylus EasyCare handpiece.

Cleaners and lubricants will damage the sealed bearings.

Note: 

 

If you do lubricate the handpiece by accident, you must continue 

lubricating the handpiece between patients.  Once the handpiece has been

lubricated even one time, it will require on-going lubrication.

IMPORTANT: 

 Use only Midwest products recommended by DENTSPLY 

to maintain your handpiece.

Stylus 

Handpiece Maintenance Instructions

After each use:

Maintenance procedures using Midwest Plus™ 
Aerosol Handpiece Cleaner/Lube (#380080) 
and XGT Nozzle (#380085)

1.

 Disconnect handpiece from coupler.

2.

 Remove bur from chuck.

3.

 Scrub and rinse under warm tap water with a soft brush or sponge 

making sure to keep head up to prevent excess water from getting into

the back of the handpiece. Make sure fiber optic port near head of 

handpiece is scrubbed well. 

4.

 While holding can of Midwest Plus Aerosol Spray upright, insert nozzle fully 

into back end of handpiece. Spray a short 1-second burst. Only if excess 

debris appears, spray another 1-second burst.  

5.

 Run handpiece, without a bur, for a minimum of 30 seconds on the hose 

or use a Midwest Air Station (#380800) and a Maintenance Coupler 

(#380090) until the handpiece reaches normal speed and all 

excess cleaner/lube has been expelled.  

6.

 Wipe off any excess lubricant on the outside of handpiece with 

a dry gauze pad or paper towel.

7.

 Insert handpiece into sterilization bag and load into autoclave.   

Midwest recommends using sterilization bags with at least one 

paper side.  Handpieces should be processed in the autoclave 

for 15 minutes at sterilization temperature of 135˚C (275˚F). For 

cassette chamber saturated steam devices, handpieces should be 

processed unbagged for 3.5 minutes at sterilization temperature 

of 135˚C (275˚F). STEAM STERILIZE ONLY. Do not use other 

means such as dry heat or chemical.

8.

 The XGT handpiece is now ready for use.

Weekly or as-needed maintenance:

If the push button gets stiff:

• Spray a 1-second burst of Midwest Plus Aerosol Spray into chuck or 

place 1-drop of Midwest Plus Handpiece Lubricant (#380130) into chuck.

To remove debris from fiber optics:

• Gently rub both ends of the fiber optic element with an alcohol soaked cotton swab.

If sluggishness occurs:

• Place handpiece, head down, in a beaker of Midwest Plus Handpiece 

Cleaner (#380140). 

• Place beaker in ultrasonic cleaner for 10 minutes.

• Clean, lubricate, and sterilize per recommended Midwest maintenance instructions.

IMPORTANT: 

 Use only Midwest products recommended by DENTSPLY 

to maintain your handpiece.

 

 

     Handpiece Maintenance Instructions

After each use:

Maintenance procedures using Midwest Plus™
Aerosol Handpiece Cleaner/Lube (#380080) and 
Midwest Nozzle (#380086)

1

. Remove bur from chuck.

2.

 Scrub and rinse under warm tap water with a soft brush or sponge making 

sure to clean both ends of the fiber optic bundles.

3.

 While holding can of Midwest Plus Aerosol Spray upright, insert nozzle 

fully into drive air tube.  The drive air tube is the shorter of the two big 

tubes at the back of the handpiece, and is a gold color on newer hand-

pieces. Spray a short 1-second burst.

4.

 Run handpiece, without a bur*, for a minimum of 30 seconds on the hose 

or use a Midwest Air Station (#380800) until the handpiece reaches normal 

speed and all excess cleaner lube has been expelled.  

5.

 Wipe off any excess cleaner lubricant on the outside of handpiece with a 

dry gauze pad or paper towel.

6.

 Insert handpiece into sterilization bag and load into autoclave.   Midwest 

recommends using sterilization bags with at least one paper side.  

Handpieces should be processed in the autoclave for 15 minutes at steril-

ization temperature of 135˚C (275˚F). For 

cassette chamber saturated steam devices, handpieces should be pro-

cessed unbagged for 3.5 minutes at sterilization temperature 

of 135˚C (275˚F).

7.

 The Tradition handpiece is now ready to use.

Maintenance procedures using Midwest Plus 

Handpiece Cleaner (#380140) and Midwest 

Plus Handpiece Lubricant (#380130)

1.

 Remove bur from chuck.

2.

 Scrub and rinse under warm tap water with a soft brush or sponge making 

sure to clean both ends of the fiber optic bundles.

3.

 Spray Midwest Plus Handpiece Cleaner liberally into chuck and into the 

drive air tube. The drive air tube is the shorter of the two big tubes at the 

back of the handpiece, and is a gold color on newer handpieces.

4.

 Run handpiece, without a bur*, for a minimum of 30 seconds on the hose 

or use a Midwest Air Station (#380800) until it reaches normal speed and 

all excess cleaner has been expelled.

5.

 Wipe off any excess cleaner on the outside of handpiece and with a dry 

gauze pad or paper towel.

6.

 Place 3 drops of Midwest Plus Handpiece Lubricant into the 

drive air tube.

7.

 Run handpiece for a minimum of 30 seconds until it reaches 

normal speed and all excess lubricant has been expelled.

8.

 Wipe off any excess lubricant on the outside of handpiece with a dry gauze 

pad or paper towel.

9.

 Insert handpiece into sterilization bag and load into autoclave.   Midwest 

recommends using sterilization bags with at least one 

paper side.  Handpieces should be processed in the autoclave for 

15 minutes at sterilization temperature of 135˚C (275˚F). For 

cassette chamber saturated steam devices, handpieces should be

  processed unbagged for 3.5 minutes at sterilizationtemperature

  of 135˚C (275˚F). STEAM STERILIZE ONLY. Do not use other means such 

  as dry heat and chemical.

 

10.

 The Tradition handpiece is now ready to use.

* Note: For Tradition Standard (wrench chuck) the bur must remain in when-

ever running/expelling or chuck will back out of the set.

Weekly or as-needed maintenance:

If the push button or power lever gets stiff:

• Place 1 drop of Midwest Plus Handpiece Lubricant (#380130) into chuck.

If sluggishness occurs:

• Place handpiece, head down, in a beaker of Midwest Plus Handpiece 

Cleaner (#380140). 

• Place beaker in ultrasonic cleaner for 10 minutes.

• Clean, lubricate and sterilize per recommended Midwest 

maintenance instructions.

IMPORTANT:

 

 Use only Midwest products recommended by DENTSPLY to main-

tain your handpiece.

 

                              Handpiece Maintenance Instructions

After each use:

Maintenance procedures using Midwest Plus™
Aerosol Handpiece Cleaner/Lube (#380080) and 
Midwest Nozzle (#380086)

1.

 Remove bur from chuck.

2.

 Scrub and rinse under warm tap water with a soft brush or sponge making 

sure to clean both ends of the fiber optic bundles.

3.

 While holding can of Midwest Plus Aerosol Spray upright, insert 

nozzle fully into drive air tube.  The drive air tube is the shorter of 

the two big tubes at the back of the handpiece, and is a gold color on newer 

handpieces. Spray a 1-second burst.

4.

 Run handpiece, without a bur, for a minimum of 30 seconds on the hose or 

use a Midwest Air Station (#380800) until the handpiece reaches normal 

speed and all excess cleaner lube has been expelled.  

5.

 Wipe off any excess cleaner lubricant on the outside of handpiece with a dry 

gauze pad or paper towel.

6. 

Insert handpiece into sterilization bag and load into autoclave.   Midwest 

recommends using sterilization bags with at least one paper side. Handpieces 

should be processed in the autoclave for 15 minutes at sterilization tempera-

ture of 135˚C (275˚F). For cassette chamber saturated steam devices, hand-

pieces should be processed unbagged for 3.5 minutes at sterilization

  temperature of 135˚C (275˚F). STEAM STERILIZE ONLY. Do not use other 

means such as dry heat or chemical.

7. 

The Quiet-air handpiece is now ready to use.

Maintenance procedures using Midwest Plus 

Handpiece Cleaner (#380140) and Midwest 

Plus Handpiece Lubricant (#380130)

1.

 Remove bur from chuck.

2.

  Scrub  and  rinse  under  warm  tap  water  with  a 

soft 

brush or sponge making sure to clean both ends of 

the 

fiber optic bundles.

3.

 Spray Midwest Plus Handpiece Cleaner liberally into chuck and into the drive 

air tube.  The drive air tube is the shorter of the two big tubes at the back of 

the handpiece, and is a gold color on newer handpieces.

4.

 Run handpiece, without a bur, for a minimum of 30 seconds on 

the hose or use a Midwest Air Station (#380800) until the handpiece 

reaches normal speed and all excess cleaner has been expelled.

5.

 Wipe off any excess cleaner on the outside of handpiece with a dry gauze 

pad or paper towel.

6.

 Place 3 drops of Midwest Plus Handpiece Lubricant into the drive air tube.

7.

 Run handpiece for a minimum of 30 seconds until it reaches 

normal speed and all excess lube has been expelled.

8. 

Wipe off any excess lubricant on the outside of handpiece with a dry 

gauze pad or paper towel.

9.

 Insert handpiece into sterilization bag and load into autoclave.   Midwest 

recommends using sterilization bags with at least one paper side.  

Handpieces should be processed in the autoclave for 15 minutes at ster-

ilization temperature of 135˚C (275˚F). For cassette chamber saturated 

steam devices, handpieces should be processed unbagged for 3.5 minutes 

at sterilization temperature of 135˚C (275˚F). STEAM STERILIZE ONLY. Do 

not use other means such as dry heat or chemical.

10.

 The Quiet-air handpiece is now ready for use.

* Note: For Quiet-air Standard (wrench chuck) the bur must remain in when-

ever running/expelling or chuck will back out of the set.

Weekly or as-needed maintenance:

If the power lever or chuck gets stiff:

• Place 1 drop of Midwest Plus Handpiece Lubricant (#380130) 

into chuck.

If sluggishness occurs:

• Place handpiece, head down, in a beaker of Midwest Plus 

Handpiece Cleaner (#380140). 

• Place beaker in ultrasonic cleaner for 10 minutes.

• Clean, lubricate and sterilize per recommended Midwest 

maintenance instructions.

IMPORTANT:

 

 

Use only Midwest products recommended by DENTSPLY to 

maintain your handpiece.

Stylus 

Standard

Stylus 

Stylus

 

EasyCare™

XGT

XGT

Tradition

Tradition

®

Quiet-air

Quiet-air

®

DRIVE AIR

TUBE

 

 

 

         

 Handpiece Maintenance Instructions

After each use:

Maintenance procedures using 
Midwest Plus™ Aerosol 
Handpiece Cleaner/Lube (#380080) 
and Stylus Nozzle (#380085)

1.

 Disconnect handpiece from coupler.

2

. Remove bur from chuck.

3.

 Scrub and rinse under warm tap water with a soft brush or 

sponge making sure to keep head up to prevent excess water 

from getting into the back of the handpiece. Make sure fiber 

optic port near head of handpiece is scrubbed well. 

4.

 While holding can of Midwest Plus Aerosol Spray upright, insert nozzle fully 

into back end of handpiece. Spray a short 1-second burst. Only if excess 

debris appears, spray another short 1-second burst.  

5.

 Run handpiece, without a bur, for a minimum of 30 seconds 

on the hose or use a Midwest Air Station (#380800) and a 

Maintenance Coupler (#380090) until the handpiece reaches 

normal speed and all excess cleaner/lube has been expelled.  

6.

 Wipe off any excess lubricant on the outside 

of handpiece with a dry gauze pad 

or paper towel.

7. 

Insert handpiece into sterilization bag and load 

into autoclave. Midwest recommends using sterilization 

bags with at least one paper side. Handpieces should 

be processed in the autoclave for 15 minutes at sterilization 

temperature (135˚C/275˚F). For cassette chamber saturated 

steam devices, handpieces should be 

processed unbagged for 3.5 

minutes at sterilization temperature 

of 135˚C (275˚F). STEAM STERILIZE 

ONLY. Do not use other means such as dry 

heat or chemical.

8.

 The Stylus handpiece is now ready to use.

Weekly or as-needed maintenance:

If the push button gets stiff:

• Spray a 1-second burst of Midwest Plus Aerosol Spray into chuck or place 

1-drop of Midwest Plus Handpiece Lubricant (#380130) into chuck.

To clean water spray holes:

• Run the cleaning wire (supplied with handpiece) back and forth in the four 

outer water spray holes and in the four inner air spray holes to loosen and 
remove any mineral build-up.

To remove debris from fiber optics:

• Gently rub both ends of the fiber optic element with 

an alcohol soaked cotton swab.

IMPORTANT:

 

 Use only Midwest products recommended by DENTSPLY                   

to maintain your handpiece.