Mid-West Delta Meter 142 Installation And Operating Instructions Download Page 1

 

 

 

BULLETIN NO. IM142/09 

 

 

 

 

 

Instrument

Mid-West

®

 

 
 

 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
INSPECTION 

 
Before installation check the nameplate on each 
instrument against the receiving paperwork and the 
intended application for correct part number, materials 
of construction, working pressure, dial range, etc.  If 
equipped with switches, check electrical rating.  Inspect 
for shipping damage and, if damaged, report it 
immediately. 
 

NOTE - Before attempting repairs contact your 
local Mid-West Representative or our factory. 

 

Failure to do so will void any warranty. 

 

PRODUCT DESCRIPTION

 

The Model 142 "Delta Meter"® is a medium range 
differential pressure instrument available as a switch, a 
gauge, or both.  See the “Part Numbering System”, 
(Fig. TBD) for available options. 
 
A flexible elastomer diaphragm and calibrated range 
spring are moved by differential pressure.  A pair of 
magnets, coupled with the diaphragm, transmit this 
motion through the wall of the pressure housing to a 
follower magnet attached to an indicating pointer. The 
rotation of the follower magnet causes the pointer to 
track the movement of the internal magnet and indicate 
the differential on the dial scale. 
 

INSTALLATION

 

The model 142  is calibrated and tested prior to 
shipment and is ready for immediate installation.  Use 
of the following installation procedures should eliminate 
potential damage and provide optimum trouble-free  
operation. 
 

 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1.  PROCESS CONNECTIONS 

 
1/4" FNPT are provided as standard however check 
the paperwork for the connections ordered.   There are 
two connections on the housing identified as "hi" and 
"lo" for high pressure and low pressure. Be sure these 
get plumbed to the proper connections on  your  
system.  Improper connection will not damage the 
instrument, but it will not function properly. Flexible 
tubing is recommended to minimize effect of possible 
vibration. 

 
2. INSTRUMENT LOCATION 

 
On liquid service the instrument should be mounted 

below

 the process connections to facilitate self-

bleeding.  On gas service it should be located 

above

 

the process connections to promote self-draining.  If 
the process contains particulates, a "pigtail" loop or 
drop leg (manometer "U-tube" configuration) in the 
tubing will minimize the possibility of it migrating into 
the instrument. 
 

3. PANEL MOUNTING 

 
Gauges with 2-½" dials can only be mounted through 
the 

rear

 of the panel.  Make the proper panel cutout as 

indicated in (Fig .TBD). Remove the (4) bezel screws.  
Insert the gauge front through the rear of the panel and 
reinstall the bezel screws through the front of the panel 
and into the gauge bezel.  Tighten the screws securely, 
alternating in a 

diagonal

 pattern. 

 
Gauges with a 3 1/2 “ dial must be mounted from the 
front of the panel.  Contact the factory for mounting 
information and dimensional data. 

MODEL 142: “DELTA METER

®

” 

Installation and Operating Instructions

 