INSPECTION 
 

Before installation carefully check the Model Number on 

each instrument against the receiving paperwork and 

the intended application. Inspect for shipping damage 

and, if damaged, report it immediately.  Verify when 
the unit is unpacked and in a vertical position that the 

pointer is on zero. If the pointer is off zero see 

TROUBLE SHOOTING  section 2 for instructions  

on zeroing the pointer. 

 

 
NOTE: BEFORE 

ATTEMPTING 

REPAIRS, 

CONTACT YOUR LOCAL MID-WEST 

REPRESENTATIVE OR OUR FACTORY.  

FAILURE TO DO SO WILL VOID ANY 

WARRANTY. 

 
 

 

 

 

 

 
 

 

 

 

 

 

 
 

 

 

 

 

 
 

 

 

 

 

 
 

 

 

 

 

 
 

 

 

 

PRODUCT DESCRIPTION 

 
The Model 116 design for cryogenic applications are 

all-metal differential pressure gauges capable of 

operating at low differential pressures for up to 500 

PSI of line pressure.    

 
The DPI is equipped with a Bi-directional Over Pressure 

Relief Valve (OPV).  When the Differential Pressure 

exceeds 130% of the range the OPV equalizes the 

pressure between the Hi and Lo sides.  Dual top and 

bottom connections are provided as standard.  The DPI 

is also equipped with a pointer zero “micro-adjust”.  If 
necessary the pointer can be re-zeroed. 

 

 

CAUTION:  
 

 

Rapid pressurization can cause severe damage 
to the sensing element. Rapid pressure change  

(either increase or decrease) is a change in 

pressure occurring fast enough to drive the 

instrument full scale in one (1) second or less.  
See installation portion of this I.O.M. for 

guidance in how to prevent rapid 

pressurization.

Model 116 DPI 

Installation and Operating Instructions

 

Mid-West

  

Instrument

BULLETIN NO.  IM116DPI/08