Micsig DP10013 User Manual Download Page 1

 

High Voltage Differential Probe 

User Manual

DP10013/DP20003

General Usage Summary

Connecting the probe: 

First, connect the probe's USB power input to a USB 

power source such as the oscilloscope's USB port and connect the probe's BNC 
connector to an oscilloscope input. Then, set the proper attenuation rate and 
connect the probe input to the device under test.

Only qualified personnel should perform service on this product.
Do not touch exposed connections and components when power is present.

Disconnecting the probe: 

First, disconnect the probe inputs from the device 

under test, and then unplug the probe output and power input.

Use proper grounding:  

To avoid an electric shock, all devices that require 

grounding must be connected to earth ground. Before making connections to 
the input or output terminals of the probe, ensure that the test instrument is 
properly grounded if necessary.

Measurement safety:

 Always be aware of the voltage rating of the probe and 

the measurement accessories you are using and of the maximum amplitude of 
the signal you intend to measure. Never apply a potential that exceeds the 
voltage ratings of the probe and/or its accessories to avoid damaging the 
product and creating a hazardous situation.

Do Not Operate in an Explosive Atmosphere.
Do Not Operate in Wet/Damp Conditions.  
Keep Product Surfaces Clean and Dry.

If an over-range condition occurs, please disconnect power and signal input 

from the probe immediately.

2)    Extending the input leads may introduce more noise during measurement. 
       If extra extension lead is necessary, please ensure the extension leads 
       are of equal length and the input signal frequency is under 10MHz. 
       Otherwise, measurement errors may occur.

Warranty

6

Micsig warrants that this probe will be free from defects in materials and 
workmanship for a period of one (1) year and accessories for a period of 
six(6)  months  from  the  date  of  shipment.  If  any  such  product  proves 

defective during this warranty period, Micsig, at its option, either will repair
the defective product without charge for parts and labor, or will provide a 

replacement  in  exchange  for  the  defective  product.  Parts,  modules 
and replacement products used by Micsig for warranty work may be new 

or reconditioned to like new performance. All replaced parts, modules and 
products become the property of Micsig.

 

Shenzhen Micsig Instruments Co., Ltd.

Add: 305 Block A, CLOU Building, Baoshen RD, North Area, Nanshan 
Science & Technology Park, Shenzhen, Guangdong, China. 518000
WEB: www.micsig.com
TEL: +86-755-88600880
E-mail: [email protected]

3)    While measuring a high frequency signal, don't touch the end of the 
        input lead with your hand or other objects. Otherwise, it may affect 
        the accuracy of the measurement.

4)

    Ensure that you use an oscilloscope with an input impedance of at 

       least

 1MΩ and bandwidth 

of at least 100MHz.

5)    Turn on the oscilloscope or externally powered instrument and let 
       the probe and equipment warm up for 20 minutes.

Best Practices

5

1)    Twisting the input leads together can help reduce noise and improve 
       the probe's high frequency response when measuring signals. Please 
       view the diagram below for an example:

6)    When a substantial change in temperature or other circumstances affect 
       the accuracy of the probe's zero point, a calibration is needed: short the 
       input terminals of the probe, then power the probe while simultaneously 
       pressing the 50X(200X) and 500X(2000X) keys for three seconds.