USER MANUAL

 

CP2100 Series AC/DC Current Probe

   

 

 

User Manual

 

 

 

Product Introduction

 

The CP2100 series current probe is a BNC type that can measure both DC and AC. With split design and compact look, 

it also can be connected to multimeters via connector. Automatic and manual zero adjustment, powered by USB, no 

need additional power supply, making measurement more convenient, often used in motor drives, industrial frequency, 

inverters, power supplies, avionics and other fields. 

 

Safety Precautions

 

 

The measurable circuit should be CAT II 600V or below 

 

Do not measure bare conductors 

 

Do not touch the measured conductor and sensor head during measurement 

 

Do not use in a humid environment 

 

Do not touch the instrument or the measured object with wet hands 

 

Please ground this product through the USB power cord 

 

Please use this product as required 

 

Specifications

 

Model

 

CP2100A

 

CP2100B

 

Bandwidth

 

DC~800KHz

 

DC~2.5MHz

 

Rise time

 

≤437.5ns

 

≤140ns

 

Range

 

10A/100A

 

Output sensitivity

 

0.1V/A

10A

;      0.01V/A

100A

 

DC accuracy (typical)

 

3%±50mA

10A

;      4%±50mA

100A

500mA~40Apk

 

15%

100A

40Apk~100Apk

 

Signal delay

 

150ns

10A

200ns

100A

 

Measuring range

 

50mA~10Apk

10A

;    1A~100Apk

100A

 

Max measurable current

 

100Apk

70.7Arms

DC+AC pk

 

200Apk-pk

70.7Arms

AC

 

Max working voltage

 

CAT III 300V CAT II 600V

 

Max float voltage

 

CAT III 300V CAT II 600V

 

Max conductor diameter

 

13mm

 

Overload indication

 

The buzzer beeps and button light flashes

 

Supplied power

 

DC 5V

 

Net weight 

 

290g

 

Package weight 

 

1000g

 

Operating temperature

 

0~40°C

 

Operating humidity

 

5%~95% (0~40

, No condensation)

 

5%~65% (40

~50

, No condensation)

 

Operation Instructions

 

1.

 

Connect BNC interface to the oscilloscope (or other instruments), then plug the USB cable to power the probe; 

2.

 

Select appropriate range according to current range, the corresponding button light will turn Green; 

3.

 

Adjust the oscilloscope settings: Input impedance 1M

Ω

; select probe to Current or display as A; Set attenuation 

ratio on corresponding channel, 100A

0.01V/A

to 100X, range 10A

0.1V/A

to 10X

 

4.

 

Press Zero button to do zero calibration, after success, the buzzer 

will “

beep

 one time; if 

beep

 three times

meaning zero calibration has failed; can also go Manual to adjustment. The external magnetic field may have 

slight influence on the DC zero position, do not move it in a large range after zero adjustment is completed; 

5.

 

Clamp the conductor under test according to the direction indicated by the jaws. Note: If the measured current 